14
IX81
Fig. 5
3
Magnification Change
(Fig. 5)
When the magnification selector knob
1
is pulled out, the magnification
will be 1.6X. When the knob is pushed in, the magnification will be 1X.
}The 1X 1.6X magnification changer lens can be replaced with a 1X 2X
magnification changer lens (IX2-CA2) by the manufacturer operation.
4
Motorized Revolving Nosepiece
# When the IX81 motorized revolving nosepiece is rotated in stand-alone
mode, automatic escape movement of an objective is not performed.
Therefore, let the motorized revolving nosepiece escape once before
turning the nosepiece.
The motorized revolving nosepiece can be rotated to switch the objec-
tive by pressing an appropriate button on the U-HSTR2 hand switch or a
button assigned by the PC.
}When the IX2-BSW software (Ver. 01.03 or later) is started up, the revolving
nosepiece escapes to avoid interference with the stage as the objective
is switched, and returns to the original position after objective changing.
# It is basically inhibited to attempt to rotate a motorized revolving
nosepiece manually.
However, it can be moved to the front and rear by hand only when an
oil immersion objective is used and bubbles are produced inside
the oil.
5
Frame Clamping Plate
}This is the module for clamping the microscope frame onto an anti-
vibration platform. The applicable anti-vibration platforms are the follow-
ing four models.
· 25 mm pitch and 50 mm pitch anti-vibration platforms.
· 1-inch pitch and 2-inch pitch anti-vibration platforms.
For the assembly procedure, see item
9
on page 58.
1