5.4 Setting the brightness mode
CLV-190 INSTRUCTION MANUAL
Ch.5
Set the brightness mode of the light source according to the endoscope to be used. The brightness
mode determines the examination light supplied to the endoscope and the method of adjusting its
intensity.
Table 5.1
WARNING
When using a fiber endoscope or a rigid endoscope without a camera head or
video converter, set the brightness mode indicator to “MANU”. Setting to “AUTO”
does not enable the automatic brightness adjustment, and the brightness may not
be adequate.
5.4
Setting the brightness mode
1
Confirm the brightness adjustment mode of the endoscope to be used according to
the Table 5.1.
2
Press the brightness mode button to select the
brightness mode. The selected brightness
mode is indicated by the brightness mode
indicators.
Figure 5.2
Brightness mode
Connected endoscopes
AUTO
EVIS EXERA III 190 series videoscope
EVIS EXERA 160, EVIS EXERA II 180 series videoscope
EVIS 100, EVIS 130, EVIS 140 series videoscope
OES fiber endoscope used in combination with the video converter
Rigid videoscope
Flexible videoscope
Rigid endoscope used in combination with a camera head (the light guide cable)
Fiber endoscope used in combination with a camera head (the light guide cable)
MANU
OES fiber endoscope
Rigid endoscope used in combination with the light guide cable (no camera head)
Fiber endoscope used in combination with the light guide cable (no camera head)
Brightness mode indicators
Brightness mode button
Summary of Contents for EVIS EXERA III
Page 2: ......
Page 6: ...iv Contents CLV 190 INSTRUCTION MANUAL...
Page 20: ...14 Summary of the Equipment Functions CLV 190 INSTRUCTION MANUAL...
Page 22: ...16 1 1 Checking the package contents list CLV 190 INSTRUCTION MANUAL Ch 1...
Page 32: ...26 2 3 Rear and side panels CLV 190 INSTRUCTION MANUAL Ch 2...
Page 98: ...92 6 4 Lamp usage indicator reset CLV 190 INSTRUCTION MANUAL Ch 6...
Page 108: ...102 7 8 Transportation CLV 190 INSTRUCTION MANUAL Ch 7...