42
4.2 Inspection workflow
CLV-190 INSTRUCTION MANUAL
Ch.4
Follow each step of the workflow for inspection of the light source before use.
4.2
Inspection workflow
1
Check the lamp ignition mode.
Section 4.3 on page 43
2
Connect an endoscope.
Section 4.4 on page 43
3
Confirm that the light source is turned ON normally.
Section 4.5 on page 46
4
Check the total accumulated operating hours of the examination
lamp.
Section 4.6 on page 48
5
Confirm that the examination lamp turns ON/OFF normally.
Section 4.7 on page 48
6
Check the brightness mode.
Section 4.8 on page 50
7
Inspect the necessary functions (dependent upon the
endoscope).
• Brightness level
Section 4.9 on page 51
• Optical-digital observation
Section 4.10 on page 56
• Transillumination
Section 4.11 on page 58
• High intensity mode
Section 4.12 on page 59
• Air and water feed
Section 4.13 on page 62
8
After inspection
Turn the examination lamp OFF if the light source will not be
used immediately.
Section 4.14 on page 64
Summary of Contents for EVIS EXERA III
Page 2: ......
Page 6: ...iv Contents CLV 190 INSTRUCTION MANUAL...
Page 20: ...14 Summary of the Equipment Functions CLV 190 INSTRUCTION MANUAL...
Page 22: ...16 1 1 Checking the package contents list CLV 190 INSTRUCTION MANUAL Ch 1...
Page 32: ...26 2 3 Rear and side panels CLV 190 INSTRUCTION MANUAL Ch 2...
Page 98: ...92 6 4 Lamp usage indicator reset CLV 190 INSTRUCTION MANUAL Ch 6...
Page 108: ...102 7 8 Transportation CLV 190 INSTRUCTION MANUAL Ch 7...