DMTA-20006-01EN, Rev. B, February 2014
Setup and Operation
21
2. Setup and Operation
This chapter contains the procedures needed to set up and operate the COBRA
scanner.
To prevent damage to the scanner components, the pipe on which you install the
scanner must be clean, and free of debris, scale, soot, or other contaminants.
2.1
Removing the Probe
When ordered as a kit (COBRA-K-XXX), the one or two probes are already installed
on the COBRA scanner at the factory. Therefore, when you want to configure the
scanner using the setup template, you need to remove the probe from the wedge.
If you need to replace a probe, and since the probe cable is permanently attached to
the probe, you will need to extract and reinsert the cable and the couplant tubes in the
protective sleeve.
To remove the probe
1.
Unscrew the probe from the wedge.
2.
Unclip the probe cable and couplant tube from the cable-management posts.