OLYMPUS Cleanliness Inspector CIX100

13  Specifications

30

System Setup Guide

13

Specifications

Move 
parameters

Specifications

Camera

SC50

Chip type

CMOS

Chip size

1/2.5 Inch

Effective area 
(width x height)

5.6 mm x 4.2 mm

Pixel size

2.2 μm x 2.2 μm

Maximum resolution

2560 x 1920 pixels

Bit depth

24 bit, 48 bit

Read-out speed

5 – 104 MHz

Interface

USB 3.0

Camera mount

Standard C-Mount

Objectives

Optical UIS2

PLAPON1.25X

NA

0.04

W.D.

5

FN

26.5

MPLFLN5X

NA

0.15

W.D.

20.0

FN

26.5

MPLFLN10X

NA

0.30

W.D.

11.0

FN

26.5

Revolving 
nosepiece

U-D6REMC
DIC slider insertion type

Reflected light 
illumination

Product name: BX3M-KMA-S

For brightfield/DIC/Polarization microscopy, built-in LED illumination

Microscope 
Frame

Reflected light 
illumination only

Product name: BX53MRF-S

Rating:
AC Adapter:
input: 100-240 V AC 50/60 Hz 0.4 A
output: 5 V DC 2.5 A
Microscope frame:
input: 5 V DC 2.5 A
Maximum power consumption:
BX53MRF-S: 5.8 W, 18.8 VA

Summary of Contents for CIX100

Page 1: ...System Setup Guide CIX100 OLYMPUS Cleanliness Inspector System Optical Microscope and Accessory...

Page 2: ...and reserves the right to make such changes without obligation to notify the purchaser In no event shall Olympus Soft Imaging Solutions GmbH be liable for any indirect special incidental or consequent...

Page 3: ...ply 10 4 Setting up the system 11 4 1 Unpacking and positioning the components 11 4 2 Connecting the cables 13 4 3 Connecting the power cord 19 5 Switching on the system 20 5 1 Inserting hardware comp...

Page 4: ...ower supply cord 32 15 Notes on conformity and disposal 35 15 1 CE Conformity Europe 35 15 2 WEEE declaration Europe 35 15 3 RoHS Conformity Europe 35 15 4 FCC conformity USA 36 15 5 For Korea only 36...

Page 5: ...performing microscopic cleanliness analyses on circular filters Undesignated use can result in injury or damage to property Olympus Soft Imaging Solutions GmbH takes no responsibility for damaged caus...

Page 6: ...d safety instructions warn of dangers when operating your system and give you useful tips Caution The warning sign and the word CAUTION indicate dangerous situations that can lead to light injuries if...

Page 7: ...r switch is on The main power switch is off Caution Damage to the eyes The microscope system contains an LED as a light source LED light can damage the eyes Do not look directly into the LED Use the l...

Page 8: ...ge Overloading the stage can hinder the movement of the stage or damage the microscope For the inspection only use membrane filters that are mounted in the filter holder Attention Device damage from o...

Page 9: ...alled on is part of a network environment note the following points Make sure that the network has appropriate virus protection The installation of operating system service packs security patches or t...

Page 10: ...or laboratory like environment Only use the system up to a maximum altitude of 2000 m The electricity supply should not fluctuate by more than 10 of the nominal voltage Pollution degree 2 IEC60664 Ins...

Page 11: ...sitioning the components Figure 2 Unpacking the components 1 Remove the components from the upper polystyrene insert 1 2 Remove the upper polystyrene insert 2 Attention Only connect the modules specif...

Page 12: ...s packaging and the monitor s packaging and remove the components and the accessories 8 Remove the bag with the desiccant and dispose of it appropriately Figure 3 System components Attention Device d...

Page 13: ...m Setup Guide 4 2 Connecting the cables Figure 4 The system s cable connections Power cord Power cord and power adapter Motor cable USB 2 0 cable DVI cable USB cable USB cable RS 232C cable 3 Z 1 X 2...

Page 14: ...Caution Always use the power cord provided by Olympus The electrical safety and electromagnetic compatibility of the device can not be guaranteed if an unsuitable power cord is used If no suitable pow...

Page 15: ...e computer s USB ports Connecting the joystick to the computer 1 Plug the joystick cable into the port on the computer Figure 6 Joystick cable and computer port Connecting the computer to the power co...

Page 16: ...p Guide 2 Plug the other end of the RS 232C cable into the computer Figure 8 RS 232 C cable and computer port 3 Connect the control box s power cord with the power adapter 4 Plug the power adapter s p...

Page 17: ...he camera s cable is located above the vertical illuminator Figure 10 Camera cable and computer port Connecting the stage to the microscope and the computer 1 Plug the motor cable 2 Y into the y port...

Page 18: ...he stage to the computer To do this plug the other end of the motor cable into the port on the computer Figure 12 Motor cable and computer port 5 Plug the power cord into the power adapter 6 Connect t...

Page 19: ...the computer to the power socket using the power cord You have now connected all of the components and can switch on the system in the next step Attention Make sure that the mains power supply where y...

Page 20: ...4 Place the particle standard device in position 2 of the multi sample holder 5 Place the filter holder with the sample in position 1 of the multi sample holder 6 Use the joystick to move the stage ba...

Page 21: ...on it are detected The system compares the results of this inspection with the known dimensions of the particle standard device If discrepancies are found a message suggesting optimization measures ap...

Page 22: ...ge you can change it in the control panel 1 Open the Control Panel dialog box You can use the search function to do this Figure 14 The Control Panel dialog box in the Microsoft Windows 10 operating sy...

Page 23: ...k the Options entry for the language you want The Language options dialog box opens 5 Click the Add an input method entry The Input method dialog box opens 6 Select the input method you want from the...

Page 24: ...Office has already been installed and activated on the system Language packs for German French Spanish Japanese and Korean are also available on the system If you want to use Microsoft Office in a dif...

Page 25: ...we recommend cleaning the particle standard device with a lint free precision wipe and some acetone or alcohol Follow the manufacturer s safety instructions for the substance that you are using 10 Tro...

Page 26: ...support olympus sis com Phone 49 251 79800 6444 fax 49 251 79800 6445 3 When both calibration processes are completed successfully perform the Check System workflow because the calibration processes i...

Page 27: ...System diagram 1 25x 5x 10x Computer and monitor System microscope Vertical illuminator Camera Camera Adapter Single port tube with lens Revolving nosepiece Joystick Stage Filterholder Particle standa...

Page 28: ...lti sample holder 1 Particle standard device with holder 1 Joystick 1 Control box 1 Power cord and power adapter for the control box 1 Connector cable Control box HP Z440 Workstation 1 Microsoft Offic...

Page 29: ...m Setup Guide Computer Monitor 1 HP Workstation 1 Power cord 1 Mouse 1 Keyboard 1 How to 1 Touchscreen monitor 1 Power cord Touchscreen Monitor 1 DVI connector cable Touchscreen Monitor HP Workstation...

Page 30: ...ount Standard C Mount Objectives Optical UIS2 PLAPON1 25X NA 0 04 W D 5 FN 26 5 MPLFLN5X NA 0 15 W D 20 0 FN 26 5 MPLFLN10X NA 0 30 W D 11 0 FN 26 5 Revolving nosepiece U D6REMC DIC slider insertion t...

Page 31: ...ed reduction gear ratio of the microscope s focus drive Motor 2 phase stepper motor Max revolutions 60 rev s Material aluminium Surface anodic coating black lacquered Stage Scan 130 x 85 Travel range...

Page 32: ...to Specifications and Certification marks for power cords below Specifications OLYMPUS is not responsible for damage caused by the use of uncertified power cords with OLYMPUS devices Voltage Rating 12...

Page 33: ...e at least one certification mark shown in table 1 If you are unable to acquire a cord tested by one of the bodies listed in table 1 in your country please use a cord that has been tested by a compara...

Page 34: ...ritish Approvals Service for Cables BASEC BASEC HAR N V KEMA KEMA KEUR HAR SEMKO AB Svenska Elektriska Materielkon troll anstalten SEMKO HAR sterreichischer Verband f r Elektrotechnik VE VE HAR Danmar...

Page 35: ...ndustrial environment requirements Immunity Complies with the requirements for industrial environments Use of the device in residential areas may cause interference 15 2 WEEE declaration Europe In acc...

Page 36: ...f not installed and used in accordance with the instructions it may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular in...

Page 37: ...OLYMPUS Cleanliness Inspector CIX100 15 Notes on conformity and disposal 37 System Setup Guide 15 6 China RoHS conformity China...

Page 38: ...48 Woerd Avenue Waltham MA 02453 U S A OLYMPUS MEDICAL SYSTEMS INDIA PRIVATE LIMITED Ground Floor Tower C SAS Tower The Medicity Complex Sector 38 Gurgaon 122001 Haryana INDIA OLYMPUS KOREA CO LTD 8F...

Reviews: