OLYMPUS Cleanliness Inspector CIX100
5 Switching on the system
21
System Setup Guide
5.2
Starting CIX ASW
1. Double click the CIX button on the desktop.
The software starts.
2. Selected a language for the user interface.
3. To do this, click the [
System Information
] > [
Languages
] button on the
software’s start page and select the required language from the list.
5.3
Check System
Perform the [
Check System
] workflow before working with the system. This
workflow checks the system and the accuracy of the calibrations. A particle
standard device, rather than a sample, is scanned and the objects on it are
detected. The system compares the results of this inspection with the
known dimensions of the particle standard device. If discrepancies are
found, a message suggesting optimization measures appears.
Prerequisite
The system has been calibrated.
For more information on the calibration processes, see the software’s
help document.
The particle standard device is clean and free of dust.
The particle standard device is in position 2 of the multi-sample holder.
1. Click the [
Check System
] button on the software’s start page.
You are taken through the [
Check System
] workflow step-by-step.
i
i
You can find more information about this workflow and using the software in
the software's help document.
Click the [
Help
] button in the software to open the help document for the
current page.