BX2 MAINTENANCE MANUAL
REPAIR PROCEDURE
- 11 -
Chapter 3.
REPAIR PROCEDURE
1. Optical adjustment
PREPARATION
Adjusting the left/right optical axis
50
60
70
position of
approx. 62mm
Objective 4X-10X
CROSS WH10X
*Adjust the interpupillary
distance to about 62mm
(See the illustration).
*Insert the cross eyepiece
into the right sleeve.
If the left/right optical axis is remarkably displaced at
checking, perform the following adjustment.
specimen
whose center is
identified
(Concentric circles etc.)
Align the specimen center with the cross center of the
WH10X eyepiece by turning the control knob of the stage.
Before adjustment
After adjustment
Cross center of
cross WH10X
specimen center
Control knob