Appendix D: Acoustic Output Information Accordance with IEC 60601-2-37
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
SSD-
5
Transducer Model:
BF-UC180F
Operating Mode:
B/M-mode
Table 101 – Acoustic output reporting table
Index label
MI
TIS
TIB
TIC
Scan
Non-scan
Non-
scan
A
aprt
1cm
2
A
aprt
>
1cm
2
Maximum Index Value
1.5
0.39
–
–
0.058
0.72
Associated
acoustic
parameters
p
r,
(MPa)
4.0
P
(mW)
12
–
0.87
14
Min of [
P
(
z
s
),
I
ta,
(
z
s
)]
(mW)
–
z
s
(cm)
–
z
bp
(cm)
–
z
b
(cm)
0.95
z
at max.
I
pi,
(cm)
0.36
d
eq
(
z
b
)
(cm)
0.22
ƒ
awf
(MHz)
6.7
6.9
–
–
6.9
6.0
Dim of
A
aprt
X
(cm)
0.67
–
–
0.67
0.67
Y
(cm)
0.26
–
–
0.26
0.26
Other
information
t
d
(
µ
sec)
0.12
prr
(Hz)
7330
p
r
at max.
I
pi
(MPa)
4.1
d
eq
at max.
I
pi
(cm)
0.22
I
pa,
at max.
MI
(W/cm
2
)
970
Operating
control
conditions
DVA%
(%)
100
95
–
–
95
100
Function
–
–
–
–
–
–
B/W Frequency Select
(MHz)
7.5MHz
7.5MHz
–
–
7.5MHz
5MHz
B/W Focus
F1
F5
–
–
F5
F6
B/W-Range
(cm)
2.0
2.0
–
–
2.0
2.0
B/W Line Density
HIGH
HIGH
–
–
HIGH
HIGH
B/W Scan Area
(%)
25
25
–
–
25
25
Doppler Image Select
n/a
n/a
n/a
n/a
n/a
n/a
Doppler Frequency Select
(MHz)
n/a
n/a
n/a
n/a
n/a
n/a
Doppler Focus
n/a
n/a
n/a
n/a
n/a
n/a
Doppler Velocity Range
(cm/sec)
n/a
n/a
n/a
n/a
n/a
n/a
Color Image Select
n/a
n/a
n/a
n/a
n/a
n/a
Color Doppler Frequency
Select
(MHz)
n/a
n/a
n/a
n/a
n/a
n/a
Flow Focus
n/a
n/a
n/a
n/a
n/a
n/a
Color Velocity Range
(m/sec)
n/a
n/a
n/a
n/a
n/a
n/a
Color Line Density
n/a
n/a
n/a
n/a
n/a
n/a
Color Average
n/a
n/a
n/a
n/a
n/a
n/a
Flow Filter
n/a
n/a
n/a
n/a
n/a
n/a
Flow Scan Area
(%)
n/a
n/a
n/a
n/a
n/a
n/a
Triplex Velocity Range
n/a
n/a
n/a
n/a
n/a
n/a
NOTE1
Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
NOTE2
Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended for transcranial or
neonatal cephalic uses.
NOTE3
Information on MI and TI need not be provided if the equipment meets both exemption clauses given in 51.2 aa) and 51.2
dd).