Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
127
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Nonsterile water rinse and alcohol flush
1.
Fill a small basin with clean water.
2.
Immerse all equipment in clean water and follow all procedures described in
“Sterile water rinse” on page 126.
3.
Fill a small basin with 70% ethyl or isopropyl alcohol.
4.
Immerse all equipment in the alcohol.
5.
Gently agitate the equipment in the alcohol.
6.
Using the 30 cm
3
(30 ml) syringe, flush the interior and recessed parts of all
equipment with alcohol to remove all air bubbles.
7.
Rub the bristles of the channel cleaning brushes and channel-opening
cleaning brush to remove all air bubbles.
8.
Attach the 30 cm
3
(30 ml) syringe to the suction cleaning adapter and flush
the adapter with 30 cm
3
(30 ml) of alcohol to remove all air bubbles.
9.
Remove the suction cleaning adapter from the alcohol and remove all
alcohol from it.
10.
Remove all equipment from the alcohol.
11.
Using a sterile, lint-free cloth, thoroughly wipe and dry all external surfaces.
12.
Dry all equipment.
13.
Store the components following the instructions given in Chapter 9, “Storage