INSTRUCTIONS FOR USE 

 

 
 
 

PERISTALTIC PUMP UNIT 

 
 
 
 

AFU-100 

 

REF:  WB950167 (100…240 V~) 

 

 

 
 
 
 

Summary of Contents for AFU-100

Page 1: ...WB135745 W10 INSTRUCTIONS FOR USE PERISTALTIC PUMP UNIT AFU 100 REF WB950167 100 240 V...

Page 2: ......

Page 3: ...to use 4 Symbols definitions and functions 4 1 Symbols and descriptions 4 1 1 Front panel 4 1 2 Rear panel 4 1 3 Packaging 4 1 4 Pump tube 4 2 Front panel 4 2 1 Buttons Switches 4 2 2 Indicators 4 2 3...

Page 4: ...3 Disposal of the unit 9 Troubleshooting 9 1 Error codes and measures 9 2 Periodic safety checks 9 3 Returning the peristaltic pump unit for repair 10 System chart 11 Specifications 11 1 Pump tube ch...

Page 5: ...d must have received appropriate training in using this peristaltic pump unit These instructions for use therefore do not explain or discuss medical procedures In addition the user must undergo an ins...

Page 6: ...nd cautions given below when handling this peristaltic pump unit This information is to be supplemented by the dangers warnings and cautions given in each chapter User related error prevention WARNING...

Page 7: ...only be used under the conditions as described in chapter 11 2 Technical data Use under other conditions may impede normal performance and or result in equipment damage Accessories WARNING Mechanical...

Page 8: ...lications DANGER Procedural hazards and complications The safety of medical procedures will be greatly enhanced by a thorough knowledge of the medical literature on the subject Study of specific infor...

Page 9: ...eristaltic pump unit can be connected to an electrosurgical unit by an interface allowing operation of the peristaltic pump unit by a special button on the electrosurgical unit s footswitch The pump r...

Page 10: ...l Power on off Caution Defibrillation proof type CF applied part cardiac application Up increase flow level Down decrease flow level Flow on off Flow activation Communication assignment with an electr...

Page 11: ...turer Date of manufacture Serial number Reference number Indication of the direction of higher volume Loudspeaker Footswitch connection Interface for electrosurgical unit connection Operating instruct...

Page 12: ...tacking limit by number including lowest box Atmospheric pressure limitation during transport and storage Relative humidity limitation during transport and storage Quantity 4 1 4 Pump tube Disposable...

Page 13: ...indicator 12 Memory 1 indicator 8 Memory 2 indicator 9 Memory 2 button 2 Footswitch assignment indicator 13 Memory 3 button 3 Memory 1 button 1 Memory 3 indicator 10 Flow activation deactivation butt...

Page 14: ...lay 14 4 2 2 Indicators 7 Error indicator This indicator illuminates red if any error occurs 8 9 10 Memory 1 2 3 indicator The corresponding indicator signals if one of the memory buttons 1 2 3 is pre...

Page 15: ...This terminal is used for potential equalization 18 Mains connector with integrated fuses This terminal serves as a connection to the mains power supply 19 Type plate The type plate contains specific...

Page 16: ...rate depends on the diameter of the inserted pump tube the used instrument including its accessories and the setting of the flow level see chapter 11 1 Pump tube characteristics chapter 7 3 Pump tube...

Page 17: ...optional item thoroughly review and understand the instructions for use provided with that item and check the compatibility with the peristaltic pump unit Specifications design and accessories are su...

Page 18: ...nform with the requirements of the IEC 60601 1 1 standard In this case observe the maximum permitted current or power loading of the multiple power socket outlet and the insulating transformer CAUTION...

Page 19: ...ng the patient in the event of a defect in the conventional protective conductor system 6 5 Connection of footswitch Before connecting confirm that the footswitch plug 22 is free of scratches and crac...

Page 20: ...l unit s footswitch activates the pump flow Align the plugs of the communication cable that the keying is facing down and insert the plugs into the communication ports of the electrosurgical unit and...

Page 21: ...ils to start up no indicators illuminates and the level display 14 is off confirm that the power cord is connected securely to a grounded wall outlet and the mains connector 18 on the peristaltic pump...

Page 22: ...tor of the pump tube left side Fig 7 1 Protective cover in opened Fig 7 2 Protective cover in opened position with inserted pump tube position with inserted pump tube front view side view WARNING Alwa...

Page 23: ...hed a warning tone can be heard The flow level can be changed while the flow is activated This peristaltic pump unit stores the last set flow level automatically By switching on the power the last flo...

Page 24: ...nction at the flow activation deactivation button 4 or the footswitch pedal 21 or the electrosurgical unit s footswitch continuous flow may cause patient injury NOTE If the peristaltic pump unit is co...

Page 25: ...fectant 3 To remove dust dirt and non patient debris wipe the peristaltic pump unit and footswitch using a soft lint free cloth moistened with 70 ethyl or isopropyl alcohol WARNING After cleaning the...

Page 26: ...d local laws and guidelines Waste electrical and electronic equipment In accordance with European Directive 2002 96 EC on waste electrical and electronic equipment WEEE the product must not be dispose...

Page 27: ...the blockage If the error occurs again please contact the technical service E25 Activation The protective cover of the pump head is opened the pump head is not connected properly Close the protective...

Page 28: ...who are authorised to carry out maintenance and repair 10 System chart The recommended combinations of equipment and accessories that can be used with this peristaltic pump unit are shown below New pr...

Page 29: ...th inner diameter of 4 8 mm and 3 2 mm CAUTION The achievable flow rate also depends on the used instrument and its accessories 11 2 Technical data Unit designation description AFU 100 peristaltic pum...

Page 30: ...m Schurter or 219 series from Littelfuse or S506 series from Bussmann Power connection line IEC 60320 1 C13 Size Width x Depth x Height 295 430 115 mm Weight 5 6 kg Volume 14588 cm Footswitch optional...

Page 31: ...r 4 8 mm REF WB920138 outer 6 4 mm inner 3 2 mm REF WB920137 Environmental conditions pump tube Normal operation Temperature 10 40 C Relative humidity 15 80 non condensing Transport Temperature 34 65...

Page 32: ...rby electronic equipment High frequency emissions CISPR 11 Class B Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 complies The peristaltic pump unit is s...

Page 33: ...ltage should comply with a typical commercial or hospital environment Surges according to IEC 61000 4 5 1 kV for differential mode 2 kV for common mode 1 kV 2 kV The quality of the power supply voltag...

Page 34: ...mitter manufacturer and d is the recommended separation distance in meters m Conducted high frequency according to IEC 61000 4 6 3 Veff 150 kHz to 80 MHz 3 V 1 2 Square root P Radiated high frequency...

Page 35: ...for use in an electromagnetic environment where the radiated high frequency disturbances are controlled The user of the peristaltic pump unit can help to prevent electromagnetic interference by maint...

Page 36: ...Jianguomenwai Avenue Chaoyang District Beijing 100022 China Fax 86 10 6569 3555 Phone 86 10 6569 3535 OLYMPUS MOSCOW LIMITED LIABILITY COMPANY OLYMPUS MOSCOW LIMITED LIABILITY COMPANY OLYMPUS MOSCOW...