3º - The tip is worn, or the ultrasonic scaler has
a defect.
3º - Contact Olsen certified assistance.
8
The ultrasonic scaler
is overheating
1º - The vibration control is not adjusted
according to tip specification.
1º - Adjust the ultrasonic scaler power control
according to the tip specification.
2º - The tip is not properly attached to the
transducer.
2º - Remove and attach the tip again using the
Torque wrench.
3º - The tip is worn, or the ultrasonic scaler has
a defect.
3º - Contact Olsen certified assistance.
9
The ultrasonic scaler
has little or no water
1º - The water tank is empty.
1º - Fill the water tank.
2º - The ultrasonic scaler water valve is closed.
2º - Open ultrasonic scaler water valve.
3º - The tip has a hydro-pneumatic system.
3º - Contact Olsen certified assistance.
10
The turbine's optical
fiber light doesn't
work
1º - The equipment is not performing the
movement.
1º - Check
Troubleshooting, item 1
.
2º -
The optical fiber’s light is switched
OFF.
2º - Turn On
the optical fiber’s light switch.
3º - The handpiece is not properly attached.
3º - Reattach the handpiece until the joint
gasket is tighter.
4º -
There’s a pneumatic or electronic problem.
4º - Contact Olsen certified assistance.
11
The electrical
micromotor doesn't
work or it's
malfunctioning
1º - The equipment is not performing the
movement.
1º - Connect the equipment to the main
electricity.
2º - Pneumatic handpieces are not working
properly.
2º - Check
Troubleshooting, item 5
.
3º -
The rotation’s direction switch is on OFF
position.
3º -
Set the rotation’s direction switch to FWD
or REV position.
4º - The speed control is adjusted to minimum
speed.
4º - Increase the micromotor's rotation speed.
5º -
There’s a pneumatic or electronic problem.
5º - Contact Olsen certified assistance.
If you are in doubt or find a problem with the equipment that is not mentioned in this chapter, stop using
the equipment immediately and contact your authorized service center or contact us by e-mail
15
- Preventive Review
In order to
extend your equipment’s lifespan, Olsen has created the extended warranty system by
performing scheduled service maintenance.
During the service maintenance, the technician will evaluate the equipment’s general maintenance
condition, as well as the monitoring of component wear and if there is a need for lubrication.
The technician may suggest the replacement of the parts with wear and tear and will provide guidance on
the daily care necessary for the proper equipment functioning.
The following tables list the items that must be checked by the technician:
PEDAL
Check of all pedal controls.
Check of the progressive foot pedal
’s
valves.
Check of the pressure regulator for instruments.
WORKING TABLE
Check of the cart castors.
Check of the pneumatic couplings pressure and rings.
Check the valves, blades, and handpieces holders.
Check of the ultrasonic scaler tips wear and tear.
Disassembly and lubrication of the 3-
way syringe’s
buttons.
CURING LIGHT
Check of the intensities and operation modes.
Check for the tips and supporting ring.
- Olsen recommends that the preventive maintenance of the items indicated in this chapter be performed