15
To select a preset program: use the Program key (8). There is 7 programs available, according the table
below:
The selected program is shown on the display (1).
To record program: select program P1 or P2, then select the
desired gear ration and adjust the speed. Press and hold the Program key
(8) until the confirmation of the system.
To adjust speed: use the adjustment keys (13) to increase or
decrease speed. Please see the table below for speed range for each
gear ration value.
The selected speed is shown on the display (5).
To select the gear ratio: use Gear Ratio key (10). There are the
following values available: 1:5, 1:1, 4:1, 10:1, 16:1 and 20:1.
The selected gear ratio is shown on the display (4).
To change the rotation direction: use the Direction key (9). he
selected direction is shown on the display (3).
To adjust the micromotor water: use the water control knob (15) on
the instruments control panel.
To access setup menu: press and hold the Setting key (12).
Use the Setting key (12) to select the desired option.
To adjust LED options: use the adjustment keys (13) to set the
parameter.
To adjust pneumatic activation mode: use the adjustment keys (13) to set the parameter. The selected
pneumatic activation method is shown on the display (6).
To adjust input pneumatic pressure: use the adjustment keys (13) to set the parameter.
Note: the micromotor system preset by the factory according to the internal equipment pressure. It is not
recommended to change the pressure parameters.
- Vórtice High Power Saliva Ejector (Optional)
The Vórtice high power saliva ejector is a device that uses
the Venturi saliva ejector system, but its volume capacity is
superior to that of the Venturi, reaching up to 385 mm/Hg.
The Vórtice high power saliva ejector may be used for
suctioning in minor surgical and prophylaxis procedures.
To enable/disable suction of the Vórtice saliva ejector (
2
): use the On/Off button (1).
The canula adapter (3) is removable for cleaning.
Set minimum activation pressure
from 0.03 to 0.10 MPa
LED
Off
LED On
without
delay
LED On
with 10s
delay
When pedal is pressed micromotor
performs the maximum speed
adjusted
Micromotor performs speed according
to the pressure on the pedal
Set maximum input pressure
from 0.20 to 0.40 MPa
Forward
Reverse