- Check List
The list below is intended to provide, in summary, important details to be checked regarding the equipment
reception, installation environment preparation, appropriate supplies conditions and its correct installation.
We recommend that the equipment’s owner accompanies the technician during this Check List so that, if
necessary, he can arrange the necessary items to guarantee the correct equipment’s installation.
1 -
Is the handpieces pressure supply adequate (2,8 to 6,0 bar)?
2 -
Is the power supply in accordance with
section 7.3
(
Electrical Installation
), with proper wiring and grounding
for the equipment?
3 -
Is there an adequate circuit breaker for equipment protection, in accordance with the
section 7.3
(
Electrical
Installation
) of this manual?
4 -
The supply voltage must be 127/220 V with a ± 10 % tolerance. Was a surge protector installed to supply the
equipment in case the voltage exceeded the tolerance?
5 -
Is the air pressure on the equipment tank between 4,5 and 7,0 bar (60 and 100 PSI)?
6 -
Does the compressor turn On and Off when performing its cycles?
7 -
Did the equipment arrived as ordered and was delivered in perfect condition?
8 -
Have the castors movements been checked?
At the verification end, make sure that the air and water connections are free of leaks.
- Olsen Accredited Technical Assistance Network
To access the Olsen certified technical assistance network for installation and maintenance or contact us
Installation should be made only by an authorized technician. Installation performed by an
unauthorized person will result in loss of warranty.
9
- Cleaning and Disinfection
The whole sanitize process must be done with the use of gloves suitable for cleaning and protection,
in addition to a mask and protective glasses, according to biosafety standards.
All equipment items mentioned in this chapter must be sanitized and sterilized (when appropriate)
prior to use.
Before starting to clean the equipment, turn off the power supply and keep the ambience clean.
This equipment does not have a water sterilization system. To ensure that the equipment
sterilization does not affect water quality, the instruments, the syringe tip and its body, must be
sterilized after each use.
- Painted and Plastic Parts
It must be cleaned with a slightly damp cloth containing only soap or mild detergent.
Never use hypochlorite or alcohol-based products.
- Sterilization by Autoclave
Prior to autoclaving, clean the items, removing any organic residue, both from the surface and internal
ducts (if any). Then carefully dry each item, including the internal ducts, if possible, with compressed air.
Do not use any type of oil on the items to perform autoclaving.
Individually pack each item, with its own sterilized packaging for the autoclaving process.
For steam autoclaving, the following values may be used (sterilization in accordance with
ISO 17665
):
a)
130º C, 2 bars, 15 min;
b)
120º C, 1 bar, 3 0 min;