26
Do not use the equipment with water pressure, compressed air or electrical voltage outside the
specifications given in Chapter 7 of this manual. Using the equipment outside the specified
conditions may cause loss of functionality. Equipment defects resulting from the equipment usage
outside of its specifications will not be covered by warranty.
Only the authorized technician can replace this equipment’s mains cable and internal fuses.
Do not replace the fuse in the electrical panel while it is switched O
n. There’s risk of electric shock.
Turn Off the On/Off switch on the electrical panel before replacing the fuse.
Do not remove the equipment covers. There is a risk of electric shock! Only the authorized technician
can perform this operation.
Do not install or use any electrical equipment over or near the
Olsen’s
Vet Cart Premium. If necessary,
the Vet Cart Premium must be checked to see if it is functioning normally in the configuration in which
it will be used.
Do not do cleaning or maintenance of the equipment while it is in use, with the patient or switched
on.
Do not perform the following procedures if it is possible to touch the patient, even if unintentionally,
during the procedure:
a)
Coupling or removal of the electrical micromotor from its coupling;
b)
Coupling or removal of the ultrasonic scaler from its coupling;
c)
Coupling or removal of high-speed optical fiber handpiece from its coupling;
d)
Fuse replacement.
- Disposal
Debris, residues and infectious materials resulting from the procedures performed on this equipment
must be deposited in biological waste duly identified and in accordance with current legislation.
For proper disposal of this equipment and its components and accessories, we recommend that it be
sent to specialized recycling companies to ensure the best destination of each component without
harm to the environment.
The draining of this equipment should not be disposed of in a common sewage system, complying
with the requirements established by the health agencies.
The disposal of this equipment should be in accordance with local laws.
- Transport and Storage
It is recommended that the equipment transportation and storage be made in its original packaging.
Transport carefully protecting equipment from falls and impacts.
Protect from moisture, rain exposure and direct contact with liquids.
Keep it sheltered from the sun.
Do not heap up more than 4 volumes.
Temperature range for transportation and storage: -10°C to +45º C.
Moisture limits for transport and storage: 20% to 70%.
Do not move or store the equipment on uneven surfaces.