5.7
Troubleshooting after calibration procedure
Problem
: There is a marker in the image continually while I do not use a
marker.
The calibration is performed with a lead marker. This is calibrated along and is
shown in the image
Remedy
: Calibrate again without marker
Problem
: There is a square or rectangle shape on each image with large
fields.
The calibration is performed with the sensor partially radiated
Remedy
: Calibrate again with collimator wide open so the complete sensor is
radiated.
Problem
: The image quality is generally lower after calibration
Calibration occurred with something between the X-ray tube and the sensor,
like a mattress
Calibration occurred within 30 minutes after system switch on. The system
needs at least 30 minutes to stabilize
Remedy
: Calibrate again, make sure nothing is between X-ray tube and
sensor and/or wait for at least 30 minutes.