User Manual: Easy DR
ODB-100-5070ENG
23
Rev.:17-04-2020
3
Operating controls
There are hardware (buttons and LED indicators) and computer controls. The
hardware controls are located on the hand control unit, on the X-ray generator
mini control panel, and on the collimator. The computer controls are the touch
screen buttons provided in the user interface of the control computer unit.
3.1
Hand control
WARNING:
Motorized movements are only allowed if operator stands
in front of and outside the working area of the Easy DR.
WARNING:
Motorized movements are only allowed if patient and
system are observed by the operator, because of squeezing
hazards.
WARNING:
Operating the Easy DR from the hand control unit is only
allowed when patient and system are observed and are
under supervision of personnel because of squeezing
hazards.
WARNING:
Due to squeezing hazard, patients shall always be placed
outside the Working area during motorized rotation of the
Easy DR.
WARNING:
Special precautions from the user has to be taken when
using the Easy DR together with a table. Squeezing hazards
exists in these situations.
The hand control unit has four push buttons they control the following
functions:
Support frame height up and down controls. With these buttons
the system can be positioned to a convenient height for the
patient
Support frame rotation controls. With these buttons the system
can be positioned from 0 to -90
˚
to accommodate multiple
radiological exams. The rotation buttons only work when the
support frame is between 850 and 1350 mm above the ground