User Manual: Easy DR
ODB-100-5070ENG
22
Rev.:17-04-2020
2.2
System safety
2.2.1
Safety switch
A safety switch is mounted underneath the detector unit. When the pressure of
an obstacle pushes on this plate, the control switch for downward and rotation
movement on the hand control unit becomes inactive. The upward movement,
however, remains possible.
2.2.2
Emergency stop button
An emergency stop button is part of the scanner unit and is located just next
to the collimator unit. In case of an emergency this button can be pressed and
it will stop any movement. To release the button, rotate it a quarter of a turn.
Before releasing the button, bring the patient into safety. Then clear the area
from obstacles.
2.2.3
Transport locking
Mobile units: Underneath the locking device bolted on the floor a switch is
mounted. When the system is locked for transport this switch is open and
functions the same way as the E-stop e.g. no movements possible.