4.9
Making a pa thorax X-ray
WARNING:
Check patients on open wounds before placing them to the
detector, if open wounds are detected, wear surgical gloves
during procedure and cleaning.
WARNING:
Motorized movements are only allowed if operator stands
in front of and outside the working area of the Easy DR.
WARNING:
Operating the Easy DR from the hand control unit is only
allowed when patient and system are observed and are
under supervision of personnel because of squeezing
hazards.
WARNING:
Due to squeezing hazard, patients shall always be placed
outside the Working area during motorized rotation of the
Easy DR.
WARNING:
Special precautions from the user has to be taken when
using the Easy DR together with a table. Squeezing hazards
exists in these situations.
•
To prevent infections, clean all parts of the equipment that are touched
by the patient. (handgrips, chin rest, front plate).
•
Place patient in the correct position.
•
Press the field illumination button on the collimator unit. The field
illumination will be projected on the patient.
•
Position the patient well in the centre (use the cross-wire of the field
illumination indicator).
•
Adjust the height using the up/down buttons on the hand control unit.