5/Troubleshooting

SYMPTOM

POSSIBLE CAUSES

POSSIBLE SOLUTION

Unit fails to provide 

vacuum at patient port 

in all modes.

Mode Selector Switch is at O(OFF) position

Move mode selector switch [34] to desired 

mode.

No Supply Vacuum

Check Hospital vacuum level (should be at 

least 500 mmHg (66.7 KPA).

Damaged/Clogged Supply side or Patient 

side fittings

Replace fittings if damaged/clogged 

Retighten/Torque if found loose

Knob fails to return when pushed

Push and rotate the knob [2] if stuck

Replace actuator [6], drive gear [5], 

Regulator Case [3], OTS assembly [10].

Knob stuck at OFF position (all the way 

counter clock wise direction)

Push and Rotate Knob [2] to clock wise 

direction

Damaged Regulator Module

Regulator Module Assembly [12]

Clogged external filter

Replace external filter

Overflow protection device is shut off

Check(Reset float)/Replace Overflow 

protection device

Loose or incorrect set-up

Set-up unit correctly without any leak.

Aspirant draw into unit

Follow your facilities procedures for handling 

contaminated products. (DO NOT SEND 

UNIT BACK TO THE MANUFACTURER)

Gauge doesn’t respond 

to changes in suction

Gauge assembly is not properly aligned.

Ensure gauge assembly [6] is properly 

aligned.

Missing/Damaged O-Rings

Replace O-Rings [9]

Gauge assembly is damaged

Replace gauge assembly [6]

Digital Unit Only:

Depleted batteries

Digital Unit Only:

Replaced Gauge [6]/Batteries

Suction level cannot be 

adjusted

Unable to push and rotate knob

Push and rotate knob [2] if stuck

Replace actuator [6], drive gear [5], 

Regulator Case [3], OTS assembly [10].

Damaged OTS assembly

Replace OTS assembly [10]

Regulator module is stuck in full OFF or full 

ON position

Rotate the knob [2] to free the Regulator 

Module Assembly [12]

Snap cap separated from regulator module 

housing.

Reassemble snap cap [28] onto regulator 

module housing [34]

Replace snap cap [28] and/or regulator 

module housing [34]

Replace Regulator Module Assembly [12]

Damaged/Missing spring inside regulator 

module assembly

Replace Regulator module assembly [12]

Damaged/Missing O-ring, Quad-ring on 

regulator module

Replace O-ring [35] and/or Quad-ring [36]

Damaged Diaphragm inside regulator 

module

Replace Diaphragm [30] or regulator module 

assembly [12]

Damaged Regulator Module Assembly

Replace Regulator module assembly [12]

Summary of Contents for Push-To-Set PTS-CVR

Page 1: ...90 100 100 200 300 400 500 600 700 kPa mm Hg L L M M H H F F MEDICAL VACUUM ANSI Analog Adult Standard 1224 Adult High 1226 Digital Adult Standard 1324 Adult High 1326 ANSI Analog Adult Standard 1225...

Page 2: ...surance Department The user of this Product shall have the sole responsibility for any malfunction which results from improper use faulty maintenance improper repair damage or alteration by anyone oth...

Page 3: ...c Unit 13 6 3 2 2 Neonatal Unit 13 6 4 Regulator Module 13 6 4 1 Disassembly 14 6 4 2 Assembly 14 6 5 Digital Gauge 14 6 5 1 Removal of Batteries 14 6 5 2 Inserting Batteries 14 7 Service Checkout Pro...

Page 4: ...ructions for Use Serial Number Manufacturer Abbreviations Used In This Manual CCW Counter clockwise decrease CW Clockwise increase CVR Continuous Vacuum Regulator MAX Full Line Vacuum inHg Inches of m...

Page 5: ...ter will help prevent this and extend the life of suction equipment 1 3 CAUTION Do not use any Loctite products or any products which contain Methacrylate Ester as an active ingredient to seal the thr...

Page 6: ...00 kPa mm Hg Analog 1230 Digital 1337 High Two Mode Continuous Adult L L M M H H F F MEDICAL VACUUM Analog 1330 Digital 1430 High Two Mode Continuous Adult mm Hg 100 200 500 700 300 400 600 MAX L L M...

Page 7: ...uum is provided In the regulated mode the vacuum source is connected through the regulator module which functions as an automatic valve Turning the suction control knob adjusts the position of the reg...

Page 8: ...l scale 1 Full scale High 0 760 mmHg 0 101 3 KPA 5 Full scale 1 Full scale Pediatric 0 160 mmHg 0 21 3 KPA 5 Full scale 1 Full scale Neonatal 0 160 mmHg 0 21 3 KPA 0 100 mmHg 0 13 3 KPA 5 Full scale 1...

Page 9: ...ine system Failure to clean and disinfect the wall outlet and pipeline system may result in damage to this equipment 4 1 1 Routine Exterior Cleaning Routine cleaning of the regulator is recommended as...

Page 10: ...ated products DO NOT SEND UNIT BACK TO THE MANUFACTURER Gauge doesn t respond to changes in suction Gauge assembly is not properly aligned Ensure gauge assembly 6 is properly aligned Missing Damaged O...

Page 11: ...e assembly Replace gauge assembly 6 Whistling buzzing noise from the unit Need to apply grease on stem on regulator module Apply grease on stem 31 Gauge responds slowly to changes in suction failed bl...

Page 12: ...est the gauge on its face Note See Section 9 1 Illustrated Parts 1 Remove the four cover screws from the back of the regulator 2 With regulator facing up carefully pull the cover assembly off the back...

Page 13: ...with a drop of removable thread locker such as Loctite 242 Repeat step 5 to verify the vacuum relief valve setting 8 Re attach the cover see Section 6 2 2 Assembly WARNING Excess Loctite may seal the...

Page 14: ...Place the O ring on the housing 5 Fold the diaphragm and place it on the piston head 6 Place the cap on the housing and press until it snaps into place 6 5 Digital Gauge Note Both batteries are positi...

Page 15: ...ith an accuracy of 1 of full range at 22 C Note All gauge needles should come to rest within the zero range bracket or return to the stop pin when no suction is being supplied Gauges which do not comp...

Page 16: ...1 mmHg 0 1 kPa for the NEO model and 8 mmHg 1 0 kPa for the high model 7 5 Maximum Vacuum Test PED NEO C A S S Models ONLY 1 Occlude the fitting patient port 2 Rotate the suction control knob fully c...

Page 17: ...ting patient and adapter probe ports with minimum inner diameters of 0 140 The unit is either directly connected to the vacuum source or connected with a tube no more than 4 in length with a minimum i...

Page 18: ...nvironmental conditions Replace lubricate and repair as needed 8 3 Repair Policy WARNING To reduce transportation personnel and or service personnel exposure to hazardous contamination clean and disin...

Page 19: ...nd Fittings CAUTION Do not use any Loctite products or any products which contain Methacrylate Ester as an active ingredient to seal the threads on the adapters probes and fittings All Ohio Medical re...

Page 20: ...acuum Relief Assembly 1 36 8700 0024 500 Quad Ring 006 Nitrile 1 18 See table below O Ring 2 014 Nitrile Buna N 1 8700 0006 000 PTS CVR Service Manual 1 19 6700 0115 400 Relief Valve Positive Pressure...

Page 21: ...21 8700 0006 000 Rev 12 08 2020 9 Ordering Information 23 15 14 13 11 1 5 4X 7 4X 6 5 4 7 7 25 2 26 3 27 1 12 16 10 8 5 18 19 20 21 22 17 24 32 31 30 29 28 35 36 34 33...

Page 22: ...hobic Filter 1 8 NPT x Tubing nipple 6730 0580 800 3 Pack 6730 0581 800 10 Pack 6730 0582 800 50 Pack Tubing 6700 0005 300 Reusable Overflow Safety Traps Description Part Number Overflow Safety Trap w...

Page 23: ...C 61000 4 2 8kV Contact 2 kV 4 kV 8 kV and 15 kV Air 8kV Contact 2 kV 4 kV 8 kV and 15 kV Air Floors should be wood concrete or ceramic tile If the floors are covered with synthetic material the relat...

Page 24: ...unity Test Level V m 385 380 390 TETRA 400 Pulse modulation 18 Hz 1 8 0 3 27 450 430 470 GMRS 460 FRS460 FM 5 kHz deviation 1 kHz sine 2 0 3 28 710 704 787 LTE Band 13 17 Pulse modulation 217 Hz 0 2 0...

Page 25: ...x is a registered trademark of S C Johnson Son Inc Kleenaseptic is a registered trademark of Metrex Research Corporation Betco is a registered trademark of Betco Corporation Cavicide is a registered t...

Reviews: