background image

15

8700-0006-000 (Rev. 12) 08/2020

7/Service Checkout Procedure

Important: 

This entire Service Checkout Procedure must 

be performed in numerical order.

7.1 Set-up

1.  Verify that there is 500 mmHg ± 10 mmHg (66.7 kPa 

± 1.3 kPa) vacuum on the supply gauge.

2. 

The supply open flow must be 80 L/min minimum.

3.  Connect the supply vacuum to the adapter/ 

probe port.

Supply 

Vacuum

500 mmHg 

(67 kPa) 

minimum and 

open flow 

of 80 L/min 

minimum

Supply Vacuum Reg. 

set to 500 mmHg + 

10 mmHg

(64 kPa + 1.3 kPa)

Supply Vacuum Gauge

0-760 mmHg (0-101.3 kPa)

Mode 

Selector 

Switch

Flowmeter

50 L/min

High Vacuum 

Calibration Gauge

760 mmHg/101.3 kPa

Occluded

Low Vacuum

Calibration Gauge

225 mmHg/30 kPa

7.2 Push-To-Set

TM

 Test

1. 

Connect the 10 L/min flowmeter to the patient port. 

2.  Move the mode selector switch to | (ON).

3.  Set 200 mmHg (26.6 kPa) on the gauge and keep 

the knob pushed in. For the pediatric unit set at 135 

mmHg (18.0 kPa), for the neonatal unit set at 100 

mmHg (13.3 kPa) and for the C.A.S.S. unit set at 140 

mmHg (18.7 kPa)

4.  Make certain leakage at the patient port connection is 

no more than 1 L/min. 

5. 

While observing the flowmeter with the knob still 

pushed in, rotate the knob. Leakage should be no 

more than 1 L/min. Release the knob and move the 

mode selector switch to O (OFF).

7.3 Gauge Test

Note:

   Analog gauges are supplied with an accuracy 

of ±5% of full scale deflection throughout their 

range. See step 4 of Section 7.4 Regulation Test.

Note: 

Digital gauges are supplied with an accuracy of 

±1% of full range at 22

°

C.

Note: 

All gauge needles should come to rest within 

the zero range bracket or return to the stop pin 

when no suction is being supplied. Gauges which 

do not comply may be out of calibration. Digital 

gauge should read “0” when no suction is applied.  

No calibration required on digital gauges.

Note:

  When checking the accuracy of the gauge on the 

analog unit, be sure that the calibration gauge 

has an accuracy of 1% of full scale deflection or 

better.

7.3.1 High Vacuum Gauges ONLY

1. 

Connect the regulator’s fitting/patient port to the high 

calibration gauge with tubing.

2.  Move the mode selector switch to | (ON).

3.  Ensure that the gauge is in agreement with the high 

vacuum calibration gauge within the ± 38 mmHg (± 

5.1 kPa) tolerance. Recommended test points are 

100, 300, and 450 mmHg (13.3, 40.0 and 60.0 kPa).

4.  Push and rotate the suction control knob fully 

clockwise  and ensure that the vacuum calibration 

gauge reads at least 450 mmHg (60.0 kPa).

5.  Three mode models ONLY:  Move the mode selector 

switch to MAX and ensure that the vacuum calibration 

gauge reads at least 450 mmHg (60.0 kPa).

6.  Push and rotate the suction control knob fully 

counter-clockwise (decrease) and verify the gauge 

reading decreases to zero.

7.3.2 Standard and PED/NEO/C.A.S.S. 

 

Vacuum Gauges ONLY

1. 

Connect the regulator’s fitting/patient port to the low 

calibration gauge with tubing.

2.  Move the mode selector switch to | (ON).

3.  Ensure that the gauge is in agreement with the 

low vacuum calibration gauge within ± 10 mmHg 

(± 1.3 kPa) for the standard model. For the PED/

NEO/C.A.S.S. models, ensure that the calibration 

gauge is within ±8 mmHg (± 1.1 kPa). Recommended 

Summary of Contents for Push-To-Set PTS-CVR

Page 1: ...90 100 100 200 300 400 500 600 700 kPa mm Hg L L M M H H F F MEDICAL VACUUM ANSI Analog Adult Standard 1224 Adult High 1226 Digital Adult Standard 1324 Adult High 1326 ANSI Analog Adult Standard 1225...

Page 2: ...surance Department The user of this Product shall have the sole responsibility for any malfunction which results from improper use faulty maintenance improper repair damage or alteration by anyone oth...

Page 3: ...c Unit 13 6 3 2 2 Neonatal Unit 13 6 4 Regulator Module 13 6 4 1 Disassembly 14 6 4 2 Assembly 14 6 5 Digital Gauge 14 6 5 1 Removal of Batteries 14 6 5 2 Inserting Batteries 14 7 Service Checkout Pro...

Page 4: ...ructions for Use Serial Number Manufacturer Abbreviations Used In This Manual CCW Counter clockwise decrease CW Clockwise increase CVR Continuous Vacuum Regulator MAX Full Line Vacuum inHg Inches of m...

Page 5: ...ter will help prevent this and extend the life of suction equipment 1 3 CAUTION Do not use any Loctite products or any products which contain Methacrylate Ester as an active ingredient to seal the thr...

Page 6: ...00 kPa mm Hg Analog 1230 Digital 1337 High Two Mode Continuous Adult L L M M H H F F MEDICAL VACUUM Analog 1330 Digital 1430 High Two Mode Continuous Adult mm Hg 100 200 500 700 300 400 600 MAX L L M...

Page 7: ...uum is provided In the regulated mode the vacuum source is connected through the regulator module which functions as an automatic valve Turning the suction control knob adjusts the position of the reg...

Page 8: ...l scale 1 Full scale High 0 760 mmHg 0 101 3 KPA 5 Full scale 1 Full scale Pediatric 0 160 mmHg 0 21 3 KPA 5 Full scale 1 Full scale Neonatal 0 160 mmHg 0 21 3 KPA 0 100 mmHg 0 13 3 KPA 5 Full scale 1...

Page 9: ...ine system Failure to clean and disinfect the wall outlet and pipeline system may result in damage to this equipment 4 1 1 Routine Exterior Cleaning Routine cleaning of the regulator is recommended as...

Page 10: ...ated products DO NOT SEND UNIT BACK TO THE MANUFACTURER Gauge doesn t respond to changes in suction Gauge assembly is not properly aligned Ensure gauge assembly 6 is properly aligned Missing Damaged O...

Page 11: ...e assembly Replace gauge assembly 6 Whistling buzzing noise from the unit Need to apply grease on stem on regulator module Apply grease on stem 31 Gauge responds slowly to changes in suction failed bl...

Page 12: ...est the gauge on its face Note See Section 9 1 Illustrated Parts 1 Remove the four cover screws from the back of the regulator 2 With regulator facing up carefully pull the cover assembly off the back...

Page 13: ...with a drop of removable thread locker such as Loctite 242 Repeat step 5 to verify the vacuum relief valve setting 8 Re attach the cover see Section 6 2 2 Assembly WARNING Excess Loctite may seal the...

Page 14: ...Place the O ring on the housing 5 Fold the diaphragm and place it on the piston head 6 Place the cap on the housing and press until it snaps into place 6 5 Digital Gauge Note Both batteries are positi...

Page 15: ...ith an accuracy of 1 of full range at 22 C Note All gauge needles should come to rest within the zero range bracket or return to the stop pin when no suction is being supplied Gauges which do not comp...

Page 16: ...1 mmHg 0 1 kPa for the NEO model and 8 mmHg 1 0 kPa for the high model 7 5 Maximum Vacuum Test PED NEO C A S S Models ONLY 1 Occlude the fitting patient port 2 Rotate the suction control knob fully c...

Page 17: ...ting patient and adapter probe ports with minimum inner diameters of 0 140 The unit is either directly connected to the vacuum source or connected with a tube no more than 4 in length with a minimum i...

Page 18: ...nvironmental conditions Replace lubricate and repair as needed 8 3 Repair Policy WARNING To reduce transportation personnel and or service personnel exposure to hazardous contamination clean and disin...

Page 19: ...nd Fittings CAUTION Do not use any Loctite products or any products which contain Methacrylate Ester as an active ingredient to seal the threads on the adapters probes and fittings All Ohio Medical re...

Page 20: ...acuum Relief Assembly 1 36 8700 0024 500 Quad Ring 006 Nitrile 1 18 See table below O Ring 2 014 Nitrile Buna N 1 8700 0006 000 PTS CVR Service Manual 1 19 6700 0115 400 Relief Valve Positive Pressure...

Page 21: ...21 8700 0006 000 Rev 12 08 2020 9 Ordering Information 23 15 14 13 11 1 5 4X 7 4X 6 5 4 7 7 25 2 26 3 27 1 12 16 10 8 5 18 19 20 21 22 17 24 32 31 30 29 28 35 36 34 33...

Page 22: ...hobic Filter 1 8 NPT x Tubing nipple 6730 0580 800 3 Pack 6730 0581 800 10 Pack 6730 0582 800 50 Pack Tubing 6700 0005 300 Reusable Overflow Safety Traps Description Part Number Overflow Safety Trap w...

Page 23: ...C 61000 4 2 8kV Contact 2 kV 4 kV 8 kV and 15 kV Air 8kV Contact 2 kV 4 kV 8 kV and 15 kV Air Floors should be wood concrete or ceramic tile If the floors are covered with synthetic material the relat...

Page 24: ...unity Test Level V m 385 380 390 TETRA 400 Pulse modulation 18 Hz 1 8 0 3 27 450 430 470 GMRS 460 FRS460 FM 5 kHz deviation 1 kHz sine 2 0 3 28 710 704 787 LTE Band 13 17 Pulse modulation 217 Hz 0 2 0...

Page 25: ...x is a registered trademark of S C Johnson Son Inc Kleenaseptic is a registered trademark of Metrex Research Corporation Betco is a registered trademark of Betco Corporation Cavicide is a registered t...

Reviews: