EN-8 STARTER 300C Portable Conductivity Meter
4
STARTER 300C OPERATION
WARNING
Do not operate the equipment in hazardous environments. The
equipment is not explosion protected.
WARNING
When using chemicals and solvents, comply with the instructions of the
chemical producer and the general lab safety rules.
4.1
Calibration
4.1.1 Selecting a standard
Always begin by calibrating your conductivity electrode. When using the STARTER 300C
conductivity meter, you must select a standard for calibration. (See 3.3)
Press and hold the
to enter setup mode and set calibration standard. Use
or
buttons to select and press
to confirm. Press
to leave setup mode.
The four predefined calibration standards are:
10 µS/cm
84 µS/cm
1413 µS/cm
12.88 mS/cm
Tables for automatic temperature compensation are programmed in the meter for each standard
(see appendix 7.1).
4.1.2 Performing a calibration
When performing calibration, Ohaus recommends using
Auto End Point Mode.
After powering
the meter on, be sure the top of the screen shows
to ensure the meter is in
Auto End point
Mode.
Auto or Manual End point Mode:
o
Press and hold
to change the
End Point Mode
.
o
When in Manual Mode, to manually reach a measurement or calibration value, you
need to press button-
when reading is stable and displays
: then the
sample reading or calibration value freezes,
blinks 3 times and freezes on
the display.
o
When in Auto End Point Mode, the meter determines when the reading is stable
then displays and locks the reading or calibration value automatically, the reading
freezes and blinks 3 times then disappears;
blinks 3 times and freezes on
the display.
Summary of Contents for STARTER 300C
Page 2: ...STARTER 300C Portable Conductivity Meter EN 1 ...
Page 4: ...STARTER 300C Portable Conductivity Meter EN 15 ...
Page 20: ......
Page 22: ...ES 2 STARTER 300C ...
Page 38: ...FR 2 STARTER 300C ...
Page 54: ......
Page 56: ...PT 2 STARTER 300C ...
Page 70: ...PT 16 STARTER 300C ...
Page 71: ...STARTER 300C PT 17 ...