2.5
Integrated stand for table top use
Starter 300C has an integrated table stand as shown, which can be pushed out to use the meter
on a lab bench or desk. To close the stand, push the middle edge of the stand toward the meter.
3
SETUP
3.1
Set temperature unit
Power the meter on by pressing
.
Press and hold
until the setup icon
appears on the display and the current
temperature unit blinks. (°C or °F)
Use
or
buttons to select the temperature unit.
Press button-
to confirm your setting.
Note: °C = 5/9 (°F -32)
Continue with MTC temperature value setting by following instructions or press button-
to go
back to the measurement screen.
3.2
Set MTC temperature value
After the temperature unit setting, the meter goes to setup the MTC temperature value setting.
Use
or
buttons to adjust the value.
Press button-
to confirm your setting.
Continue with calibration standard setting by following instructions or press button-
to go
back to the measurement screen.