EN-10 SCOUT
TM
SJX SERIES BALANCES
When the Security Switch is in locked position, see figure below, the menu settings are affected
as follows:
External calibration function is hidden
The readability will be reduced by a factor of 10 or brackets will be displayed around the
last digit, depending on the model
Units are locked to the current setting
Stable range locked to 1d
AZT locked to 0.5d
Zero range forced to 2%
Filter Level is locked to current setting.
Stable only in Print menu locked to ON
Continuous/Interval print cannot be selected (SJX…N/E models only)
Security switch:
Unlocked
Locked
Figure 4-1. Security switch.
4.2 Calibration Menu
Enter this menu to perform calibrations.
Notes:
* These sub menus are not available in SJX/E models. The settings vary with model.
Internal Calibration
[
In.CAL
]
Initiates an internal calibration procedure.
Automatic Calibration
[
A.CAL
]
Initiates an automatic calibration procedure.
OFF
= disabled
ON
= enabled
Calibration Adjustment
[
CAL.ADJ
]
Set the calibration adjustment factor for the internal calibration mass.
-
100…0…100
Span
[
SpaN
]
Initiates a span calibration procedure (zero and span)
Lin
[
LIN
]
Initiates a linearity calibration procedure (zero, mid-point and span).
End Cal
[
End
]
Advance to the next menu or return to the top of the current menu.
Internal Calibration* Perform
Automatic Calibration* Off, On
Calibration Adjustment* -
100…
0
…100
Span:
Perform
Linearity:
Perform
End Calibration:
Exit menu
*Bold
always represents factory default value.