background image

 

EN-16                                                                                                 SCOUT

TM

 SJX SERIES BALANCES                                                                                                 

 

 
5.3 Service Information  

If the troubleshooting section does not resolve or describe your problem, contact your 
authorized OHAUS service agent. For service assistance or technical support in the United 
States call toll-free 1-800-672-7722 ext. 7852 between 8:00 AM and 5:00 PM EST. An OHAUS 
product service specialist will be available to provide assistance. Outside the USA, please visit 
our web site, 

www.ohaus.com

 to locate the OHAUS office nearest you. 

 

5.4 Accessories

 

TABLE 5-2. ACCESSORIES 

DESCRIPTION 

Item Number 

 

DESCRIPTION 

Item Number 

RS232 kit 

30268982 

 

Stacking Kit, x1 

30268988 

USB Host kit 

30268983 

 

Specific Gravity kit 

30269020 

USB Device Kit 

30268984 

 

Auxiliary Display Kit 

30269019 

Bluetooth Kit* 

30268985 

 

Carrying Case 

30269021 

Ethernet Kit 

30268986 

 

In-Use Cover 

30269022 

Stacking Kit, x6 

30268987 

 

Printers and Cables 

Contact OHAUS 

Note: * Bluetooth kit is only available in certain regions according to the local regulations. 

 

6. LEGAL FOR TRADE

 

When the balance is used in trade or a legally controlled application it must be set up, verified 
and sealed in accordance with local weights and measures regulations. It is the responsibility of 
the purchaser to ensure that all pertinent legal requirements are met. 
 

6.1 Settings

 

Before verification and sealing, perform the following steps in Order: 

1.    Verify that the menu settings meet the local weights and measures regulations. 
2.    Units menu should be reviewed. Verify the units turned 

On

 meet the local weights and 

 

measures regulations. 

3.    Perform a calibration as explained in Section 4.2. 
3.    Set the position of the Security Switch to the locked position. 

 

6.2 Sealing

 

After the Balance has been verified, it must be sealed to prevent undetected access to the 
legally controlled settings. Before sealing the device, ensure that the security switch is in the 
Locked position. 
If using a wire seal, pass the sealing wire through the holes in the security switch and Bottom 
Housing as shown. 
If using a paper seal, place the seal over the security switch and Bottom Housing as shown. 

 

 

 

Unlocked 

Locked with Paper Seal 

Locked with Wire Seal 

 
 

 

Summary of Contents for Scout SJX Series

Page 1: ...Scout TM SJX Series Balances Instruction Manual ...

Page 2: ...EN 2 SCOUTTM SJX SERIES BALANCES ...

Page 3: ... or servicing this equipment Failure to comply with these warnings could result in personal injury and or property damage Retain all instructions for future reference Verify that the local AC power supply is within the input voltage range printed on the AC adapter s data label Only connect the AC adapter to a compatible grounded electrical outlet Do not position the scale such that it is difficult...

Page 4: ...curity slot is provided at the rear of the balance allowing the balance to be secured by an optional cable and lock accessory 2 2 Selecting the Location For best performance the Scout balance should be used in a clean stable environment Do not use the balance in environments with excessive drafts with rapid temperature changes near magnetic fields or near equipment that generates magnetic fields o...

Page 5: ...alibration When the Balance is first installed and when it is moved to another location it must be calibrated to ensure accurate weighing results For SJX E models the balances must be calibrated by using the external calibration weights For SJX models the balances have built in internal calibration system which can calibrate the balance without the need for external calibration masses Alternately ...

Page 6: ... Turns the balance Off None Units Changes the weighing unit None Mode Allows changing the application mode None Menu Enter the User menu View the preset Tare value Menu Function Short Press Yes Accepts the current setting on the display No Advances to the next menu or menu item Rejects the current setting on the display and advances to the next available setting Back Moves Back to previous menu it...

Page 7: ... hold the On Zero Off button until OFF is displayed 3 3 Calibration Operation When the balance is operated for the first time a span calibration is recommended to ensure accurate weighing results Before performing the calibration be sure to have the appropriate calibration weights Ensure that the Security switch is set to unlocked position SJX E Press and hold Menu until mMeNU Menu is displayed Wh...

Page 8: ...ored APW Proceed to step 7 4 Press Yes to establish an APW 5 The balance will then display the sample size i e PUt 10 Press No or Back to toggle the choices 5 10 20 50 or 100 6 Put the indicated number of pieces on the pan then press Yes to calculate the APW The display shows the piece count 7 Add additional pieces until the desired count is reached 8 To clear the stored APW press and hold Mode un...

Page 9: ...under the target weight range the under icon will light If the sample is within the target weight range the accept symbol will light If the sample is over the target weight range over icon will light Note Press Function to view the low and high check limits Positive Check Positive check is used to determine when the material added to the balance is within the target range In this case the UNDER an...

Page 10: ...s and hold Mode until total is displayed When Clr tot is displayed press Yes 3 9 Density Mode This mode allows the user to calculate the specific gravity of a sample when using the density kit Density mode has been initiated when G and the current unit is displayed The sample is weighed in air suspended from the hook and then weighed in water The formula is 𝑑𝑟𝑦 𝑤𝑒𝑖𝑔ℎ𝑡 𝑑𝑟𝑦 𝑤𝑒𝑖𝑔ℎ𝑡 𝑤𝑒𝑡 𝑤𝑒𝑖𝑔ℎ𝑡 Prepara...

Page 11: ... to begin The rEAdY Ready will be displayed 4 Place samples to be weighed on the pan 5 The stable value will be held on the display the Hold icon H will blink 4 MENU SETTINGS The User Menu allows the customizing of balance settings Note Additional Sub Menus may be available if Interface Options are installed See Interface User Manual for the additional setting information 4 1 Menu Navigation User ...

Page 12: ...ration Menu Enter this menu to perform calibrations Notes These sub menus are not available in SJX E models The settings vary with model Internal Calibration In CAL Initiates an internal calibration procedure Automatic Calibration A CAL Initiates an automatic calibration procedure OFF disabled ON enabled Calibration Adjustment CAL ADJ Set the calibration adjustment factor for the internal calibrat...

Page 13: ...isplay will maintain zero until a change of 3 divisions per second has been exceeded Stable Range Stable Set the amount the reading can vary while the stability symbol remains on 0 5d 0 5 balance division 1d 1 balance division 2d 2 balance division 5d 5 balance division Back Light b Light Sets backlight functionality OFF always off ON always on AUTO turns on when a button is pressed or the display...

Page 14: ...ates modes so they will be available for use with the Mode button Reset reset Reset the Mode menu to factory defaults NO not reset YES reset Weigh wWeIGH Set the status OFF disabled ON enabled Count Count Set the status OFF disabled ON enabled Percent Percnt Set the status OFF disabled ON enabled Check CHECk Set the status OFF disabled ON enabled Totalization totAL Set the status OFF disabled ON e...

Page 15: ...being held on the display the Hold icon will blink and the displayed weight will not change Advance to the next menu or return to the top of the current menu End Mode End 4 5 Unit Menu This menu activates units so they will be accessible with the Units button The units in the menu must be turned on to be active Note Available units vary by model and local regulations g off on kg off on ct off on o...

Page 16: ...h below The balance can be supported using lab jacks or any other convenient method Ensure the balance is level and secure and that the transportation lock has been released Power on the Balance then use a string or wire to attach items to be weighed Connecting the Interface Use an optional interface connectivity kit to connect the balance either to a computer printer or OHAUS auxiliary display Be...

Page 17: ...ssible Cause Cannot turn on No power to balance Poor accuracy Improper calibration Unstable environment Cannot calibrate Security switch in locked position Err 8 1 Weight exceeds power on zero range Err 8 2 Weight below power on zero range Err 8 3 Over load weight exceeds rated capacity Err 8 4 Under load pan removed Err 8 5 Tare out of range Err 8 6 Displayed value exceeds 999999 possible in Tota...

Page 18: ...regulations 6 LEGAL FOR TRADE When the balance is used in trade or a legally controlled application it must be set up verified and sealed in accordance with local weights and measures regulations It is the responsibility of the purchaser to ensure that all pertinent legal requirements are met 6 1 Settings Before verification and sealing perform the following steps in Order 1 Verify that the menu s...

Page 19: ...ty x Readability lb 1 36690 x 0 00005 3 30690 x 0 00005 1 3670 x 0 0005 7 0550 x 0 0005 13 6685 x 0 0005 18 080 x 0 005 Capacity x Readability grn 987 66 x 0 02 4938 4 x 0 2 9568 0 x 0 2 23148 4 x 0 2 9570 x 2 49380 x 2 95680 x 2 126540 x 20 Stabilization Time s 1 5 1 1 5 1 Construction ABS plastic housing with 304 stainless steel SST pan Draftshield Yes No Calibration SJX E User selectable extern...

Page 20: ...ction Weighing Units g kg ct Application Modes Weighing Parts Counting Percent Weighing Check Weighing Totalization Density Determination Display Hold Typical Battery Life 80 hours Specified Temperature Range 10 C to 30 C at 10 to 80 relative humidity non condensing Storage Conditions 20 C to 55 C at 10 to 90 relative humidity non condensing Communication RS232 USB Host USB Device Ethernet or Blue...

Page 21: ...Weighing Check Weighing Totalization Density Determination Display Hold Typical Battery Life 80 hours Specified Temperature Range Class II 10 C 50 F to 30 C 86 F at 10 to 80 relative humidity non condensing Class III 10 C 50 F to 40 C 104 F at 10 to 80 relative humidity non condensing Storage Conditions 20 C 4 F to 55 C 131 F at 10 to 90 relative humidity non condensing Communication RS232 USB Hos...

Page 22: ...d service organization of the authorized representative within the EC or by the national weights and measures W M authorities The first stage of the initial verification has been carried out at the manufacturer s work It comprises all tests according to the adopted European standard EN45501 1992 paragraph 8 2 2 If national regulations limit the validity period of the verification the user of the w...

Page 23: ...S 003 ISO 9001 Registration In 1994 OHAUS Corporation USA was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International BVQI confirming that the OHAUS quality management system is compliant with the ISO 9001 standard s requirements On June 21 2012 OHAUS Corporation USA was re registered to the ISO 9001 2008 standard LIMITED WARRANTY OHAUS products are warranted agai...

Page 24: ...ededor del mundo Avec des bureaux dans le monde entier Weltweite Geshä ftsstellen Con uffici in tutto il mondo www ohaus com 30269040 P N 30269040 C 2016 OHAUS Corporation all rights reserved todos los derechos reservados tous droits réservés Alle Rechte vorbehalten tutti i diritti riservati Printed in China Impreso en la China Impriméen Chine Gedruckt in China Stampato in Cina ...

Reviews: