MB27 EN-9
6.
TECHNICAL DATA
Admissible Ambient Conditions
Location:
Indoor use only
Temperature:
10°C to 40°C
Relative humidity:
15 % to 80 % at 30°C non-condensing
Warm up time:
At least 15 minutes after connecting the Analyzer to the AC supply;
when switched from standby mode the Analyzer is ready for immediate use.
Height above sea level: Up to 2000 m
Power Input:
100VAC - 120VAC, 3A, 50/60Hz or 200VAC -240VAC, 3A, 50/60Hz
Voltage fluctuations:
-15% +10%
Power load:
250W
Power line fuse:
1 piece, 5 x 20mm, 2.5A 250V
Protected against dust and water, Pollution degree: 2, Installation category: Class II
Specifications
MODEL
MB27
Capacity
90g
Readability (% requires >10g sample)
0.001g ; 0.01%
Temperature Settings
50
°C
to 160
°
C (5
°
increments)
Heat Source
Halogen
Calibration - Weight
50g
Calibration - Temperature
OHAUS Temperature Calibration Kit
Pan Size
90mm diameter
Unit Dimensions (DxWxH)
11.0x6.5x5.0 in / 28.0x16.5x12.7 cm
Unit Weight
4.6 lb / 2.1 kg
Shipping Dimensions (LxWxH)
20.2x15.5x13.5 in / 51x40x35 cm
Shipping Weight
10 lb / 4.5 kg
Communication
RS232 Pin Connections
Female DB9 connector
Pin 2: Analyzer transmit line (TxD)
Pin 3: Analyzer receive line (RxD)
Pin 4 and 6: Internally connected
Pin 5: Ground signal (GND)
Pin 7: Clear to send (hardware handshake) (CTS)
Pin 8: Request to send (hardware handshake) (RTS)
RS232 Data Settings
Baud Rate(default):
2400
Data Bits:
7
Parity:
N
Stop Bits:
2
Flow Control:
Xon/Xoff
Baud Rate:
9600
Data Bits:
8
Parity:
N
Stop Bits:
1
Flow Control:
Xon/Xoff
CTS
RTS
5
9
1
6
RxD
TxD
GND
Summary of Contents for MB27
Page 2: ......
Page 14: ...EN 12 MB27...
Page 26: ...ES 12 MB27...
Page 38: ...FR 12 MB27...
Page 50: ...DE 12 MB27...
Page 62: ...IT 12 MB27...
Page 63: ......