EN-10 MB27
RS232 Commands
The RS232 Interface allows a computer to control the Analyzer, as well as to receive data such as displayed weight.
Command
Function
ON
Turns Analyzer ON
OFF
Turns Analyzer OFF
T
Same as pressing
Tare
U
Same as pressing
%g
START
begins a test
STOP
ends a test
P
Same as pressing
xP
Interval Print x = Print Interval (1-3600 sec)
PSN
Print Serial Number
PV
Print software version
?
Print Header
H
Header on or off
RS
Print current RS232 settings
RS:2400,7,N,2,X
Change RS232 setting (The current RS232 setting can be displayed on the LCD by pressing
for 2 seconds while in standby mode.)
Baud:1200,
2400
,4800,9600,19200
Data Bits:
7
or 8
Parity:
N
= none, O = odd, E = even
Stop Bits: 1 or
2
Handshake:
X
= xon/xoff (software), R = RTS-CTS (hardware), N = none
The Analyzer will return “ES” for invalid commands.
All communication uses standard ASCII format.
Sent commands must terminate with a Line Feed or Carriage Return Line Feed (CRLF).
RS232 Output
^^^^^^^^^^^^^^^^^^^^^^^^
MOISTURE DETERMINATION
OHAUS MB2x SN #########
Switchoff Mode TIMED 10:00
Drying Temp
100C
Result Units
%Moisture
00:10 0.0%MC
00:20 0.0%MC
Elapsed Time
00:02:21
Initial Weight
8.560 g
Final Weigh
8.555 g
Final Result
0.0%MC
Header on
Header on
Header on
Header on
Header on
If
pressed or print interval set.
If
pressed or print interval set.
Printed at successful end of test.
Printed at successful end of test.
Printed at successful end of test.
Printed at successful end of test.
^^^^^^^^^^^^^^^^^^^^^^^^