3-10
CHAPTER 3 SCALE TESTING AND CALIBRATION
3.6.3 Off-Center Load Test
The Off-Center Load Test is used to determine whether displayed weight values will be affected by moving
the sample to different areas of the weighing surface. See Figure 3-4.
This test must be performed using a single mass or the masses must be stacked.
1. Zero the scale.
2. Place a mass equal to 1/3 of the full-scale capacity in the center of the weighing surface. Press
the
ZERO
button to return the reading to zero.
3. Slide the mass halfway between the center and the rear edge of the platform, and record the reading.
4. Slide the mass to the center of the weighing surface and press the
ZERO
button to return the reading
to zero.
5. Slide the mass halfway between the center and the left edge of the platform, and record the reading.
6. Slide the mass to the center of the weighing surface and press the
ZERO
button to return the reading
to zero.
7. Slide the mass halfway between the center and the right edge of the platform, and record the
reading.
8. Slide the mass to the center of the weighing surface and press the
ZERO
button to return the reading
to zero.
9. Move the mass halfway between the center and the front edge of the platform, and record the
reading.
10. Maximum reading should be less than or equal to the OCL specifications for the scale.
Reading toward Rear
____________
Reading toward Left
____________
Reading toward Right
____________
Reading toward Front
____________
Maximum Reading
____________
Figure 3-4. Off-Center Load Test Weight Locations.