CHAPTER 3 REPAIR PROCEDURES
EB Series Scales Service Manual
3-13
Ohaus Corporation www.ohaus.com
3.5.1 Overload Protection Stop Adjustment
continued
3. Add the mass specified in Table 3-2. The mass should be placed on the center of the
Pan. With the specified mass on the Pan, all of the overload (down) stops should hit
the bottom Load Cell Frame.
4.
Tighten the nut on the down stop screws so they will not move and recheck.
TABLE 3-2. LOAD CELL OVERLOAD PROTECTION STOP GAP SETTINGS.
SCALE CAPACITY
OVERLOAD GAP SETTING
MASS
3 kg
6 kg
15 kg
30 kg
0.6mm
0.8mm
1.0mm
1.6mm
2
4.5kg
2
9
2.
kg
22.5kg
45
.5
kg
5. Reassemble the scale.
6. Calibrate the scale using Appendix B instructions. Check all operations.