CHAPTER 1 GETTING STARTED
Ohaus Corporation www.ohaus.com
1-4
EB Series Scales Service Manual
1.6 SCALE SETTINGS
The EB scale contains four user-selectable scale settings (Setup mode). These are as follows:
Scale Increment, Backlight, Zero Tracking Range, and Filtering. A description of each setting is
explained below.
1. To enter the setup mode, press and then release
COUNT
and
M+
at the same time to
enter into the user-selectable scale settings.
2. In the Setup mode, press
▲ to step through available settings.
3. Press
Enter
to accept the displayed setting and proceed to the next parameter.
4. Press
Exit
to proceed to the next parameter without saving any changes.
Table 1-3. MENU STRUCTURE
inC 01
EL AU
AC 2d
Fil 1
AU
ON
OFF
0.5d
1d
2d
4d
0
1
2
3
(Levels)
(d=scale division)
(mode-dependent
settings)
inC 01
EL AU
AC 2d
FiL 1
Scale increment
- Sets the displayed scale increment (also known as readability or
graduation). Selectable setup values will be model dependent and are equivalent
to within 30,000 to 3,000 scale divisions.
Backlight
- Sets the activation mode of the backlight.
The following settings are available:
AU
(Auto-on with items > 9d placed on the
Platform or any button pressed; turns off after 5 seconds of inactivity),
On
,
oFF
.
Zero Display Range
- Sets the range in which the Zero reading is maintained.
The following settings are available:
0.5d, 1d, 2d, 4d
(d= scale division).
Filtering
- Sets the level at which the Stable indication turns on; the higher the
setting, the faster the stabilization time.
The following settings are available:
0, 1, 2, 3
(levels).