CHAPTER 2 DIAGNOSIS
Ohaus Corporation www.ohaus.com
2-4
EB Series Scales Series Service Manual
TABLE 2-5.
BALANCE READING INCORRECT OR UNSTABLE.
SYMPTOM
PROBABLE CAUSE
REMEDY
Balance reads
incorrectly.
Balance reading
unstable.
Needs calibration.
Load Cell is damaged.
Main PC Board defective.
Perform Service Calibration in
Appendix B. If calibration fails, refer to
Table 2-6.
Test the Load Cell (Section 2.7). If
defective replace it (Section 3.4).
After repair, proceed with Performance
Tests in Chapter 4.
Test the Main PC Board (Section 2.8).
If defective, replace it (Section 3.3).
After repair, proceed with Performance
Tests in Chapter 4.
TABLE 2-6.
CANNOT CALIBRATE THE SCALE.
SYMPTOM
PROBABLE CAUSE
REMEDY
Scale can be turned
on but will not
calibrate.
Incorrect weights.
Load cell assembly
defective.
Main PC Board defective
Verify that proper weights are used.
Test the Load Cell (Section 2.7). If
defective replace it (Section 3.4).
After repair, proceed with Performance
Tests in Chapter 4.
Test the Main PC Board (Section 2.8).
If defective, replace it (Section 3.3).
After repair, proceed with Performance
Tests in Chapter 4.