Ohaus Corporation www.ohaus.com
20
Defender
™ Series Bases Service Manual
4.5.3 Full Load Test
1.
After the Off Center Load Test has been passed, test the Base’s full capacity according to
the Specification Table 1-1. The Base must meet all specifications as listed.
2. If the Base fails the Full Load Test, check and set the Down Stops.
(See Section 3.3.)
ATTENTION:
In all cases where a part is replaced, the base must be thoroughly checked after
the replacement is made. The base
MUST
meet the parameters of all applicable
specifications in this manual.