™ Series Bases Service Manual
4.3
Calibration with an External Indicator
1. Install the Load Cell in position. (See Section 3.2.)
2. Connect the excitation, signal shield and sensing wires (if provided), to the Indicator. Calibrate
the Indicator to read the load in the units required, for example, kilograms or grams. (Consult
the Indicator instruction manual for calibration procedures.) For calibration mass weights, see
Table 1-1, in Chapter 1.
3. Check that the Indicator is reading correctly over a range of values.