Ohaus Corporation www.ohaus.com
19
Defender
™ Series Bases Service Manual
4.5.2 Adjusting Off Center Load
If the Off Center Load (Shift) is excessive, perform adjustment as follows:
Figure 4-2. Off Center Load adjustment points.
1. Place the test weight in the center of the Platform.
2. Tare the indicator.
3. Move the weight to point A and record the reading.
4. Move the weight to point B and record the reading.
5. Move the weight to point C and record the reading.
6. Move the weight to point D and record the reading.
7. If the reading at point A is negative, file at points 1 and 4 AT AN ANGLE.
8. If the reading at point B is negative, file at points 1 and 2 STRAIGHT ACROSS.
9. If the reading at point C is negative, file at points 2 and 3 AT AN ANGLE.
10. If the reading at point D is negative, file at points 3 and 4 STRAIGHT ACROSS.
Note:
It is not recommended that you try to adjust more than
–5 counts if the beam has been
filed already. If the beam has not been filed previously, you can adjust
–10 counts. Remember,
when filing you are weakening the beam. File a little at a time.
Top view of Load Cell
Side view of Load Cell
1
2
3
4