EN-8 BW SERIES COMPACT BENCH SCALE
6. CALIBRATION AND SEALING
6.1 Calibration
For best results, calbrate the scale at regular ntervals. Temperature changes,
geographc gravty varatons, alttude changes and abuse are few reasons why a
scale may need recalbraton.
Note
: The followng procedure requres the use of hand tools and calbraton
weghts. If unable to perform the calbraton, contact an authorzed Ohaus servce
organzaton.
1. Remove the battery cover to access the two calbraton pns located n
a slot at the bottom of the scale, then short the calbraton pns (a small
slotted screwdrver may be used).
Note
: When accessng the bottom of the scale, avod placng the scale
upsde down on the pan or sub-platform. Place the scale on ts sde.
2. “
SEtUP
” then the frst parameter “
CtY
” s dsplayed.
3. Press
ON/OFF-Yes
successvely to move through the succeedng
parameters: “dEF NO” “S1”, “S2”, “S3”, “S4”, “S5”, “S6”, “dSP”, “S8”, “S9”,
“S10”, “S11”, “S12”, “bAt”, “GEO”.
Note
: Do not change the settngs of these parameters. These are servce
setup parameters ntended for authorzed servcng use only.
4. After “GEO”, “
CAL NO
” s dsplayed. Press
TARE-No
to swtch the settng
to “
CAL YES
”.
5. “
CAL YES
” s dsplayed. Press
ON/OFF-Yes
to enter calbraton sequence.
Calbraton Swtch