3-26
CHAPTER 3 REPAIR PROCEDURES
3.2.11
Unmounting the Load Cell from the Base Plate (Cont.)
Base Plate Installation
1. Make sure the mating surfaces on the Base plate and the Load Cell as shown in Figure
3-37 are clean.
2. Position the Load Cell on the base and install the 2 mounting screws. Do not tighten fully.
3. Looking down at the top of the load Cell, align it with the Base Plate as shown in Figure
3-38. Now carefully tighten the 2 mounting screws and recheck alignment.
Figure 3-38. Load Cell Alignment to Base Plate.
Alignment Area
4. Install the Weigh Below Hook and screw at the bottom of the Base Plate.
3.2.12 Changing the Load Cell PC Board
1. Remove Load Cell from Mounting Plate, refer to paragraph 3.2.11.
2. Unsolder Temperature Sensor board cable, see Figure 3-39.
Unsolder Wires
Temperature Sensor
Board
Figure 3-39. Temperature Sensor Board Unsoldering Location.