1-3
CHAPTER 1 INTRODUCTION
TABLE 1-1. ADVENTURER AV MODELS CALIBRATION POINTS
WEIGHT CLASS
MODEL
CAPACITY
SPAN LINEARITY
ALT SPAN
ASTM OIML
CAL POINTS (g) CAL POINTS (g) CAL POINT (g)
2
F1
AV53
51g
30, 40, 50
25 / 50
30, 40
1
E2
AV64 & C
65g
40, 50, 60
20 / 50
40, 60
1
E2
AV114 & C
110g
50,100
50 / 100
50
1
E2
AV264 & C
260g
150, 250
150 / 250
150
1
E2
AV213 & C
210g
100, 200
100 / 200
100
1
E2
AV313 & C
310g
150, 200, 250, 300 150 / 300
150, 200, 250
1
E2
AV413 & C
410g
200, 300, 400
200 / 400
200, 300
2
F1
AV212 & C
210g
50, 100, 150, 200
100 / 200
50, 100, 150
2
F1
AV412 & C
410g
100, 200, 300, 400 200 / 400
100, 300, 400
2
F1
AV812 & C
810g
200, 300, 400, 500 400 / 800
200, 300, 400,
600, 700, 800
600, 700, 800
1
E2
AV2102 & C
2100g 1000, 2000
1000 / 2000
1000
1
E2
AV3102 & C
3100g
1000, 2000, 3000
1500 / 3000
1000, 2000
1
E2
AV4102 & C
4100g
2000, 3000, 4000 2000 / 4000
2000, 3000
2
F1
AV2101 & C
2100g
500, 1000, 1500, 2000 1000 / 2000
500, 1000, 1500
2
F1
AV4101 & C
4100g
1000, 2000, 3000, 4000 2000 / 4000
1000, 3000, 4000
2
F1
AV8101 & C
8100g
2000, 3000, 4000, 4000 / 8000 2000,3000,4000,
5000, 6000, 7000,
6000, 7000, 8000
8000
1.4 TEST MASSES REQUIRED
The masses required to test the Adventurer AV Models and Adventurer AS Models balances must
meet or exceed the requirements of ASTM / OIML Class specified. The calibration points for the
Adventurer AV Models are listed in Table 1-1 and the calibration points for the Adventurer AS Models are
listed in Table 1-2. Use the minimum number of masses to total the calibration point value. Bolded values
are default settings.