CHAPTER 2 DIAGNOSIS
SYMPTOM
Balance can be turned on
but will not calibrate using
internal calibration.
PROBABLE CAUSE
Main PC Board is defec-
tive.
Load cell motor assem-
bly inoperative (strain
gauge only).
Load cell motor assem-
bly inoperative ( MFR
only).
REMEDY
Verify that all other functions are opera-
tional. If fuctions are OK, continue with the
procedure. If certain functions fail or can-
not be set, the Main PC board is defective.
Perform ->Repair Procedures 3.1.3 or
3.2.3
Remove the cover from the load cell.
Check the operation if any of the Internal
drive motor. The cover has to be con-
nected to the base housing and power
applied. Notice if any obstructions or belt
drive is properly set in place. Cycle the
internal calibration on and off several times,
if the motor drive fails to respond, it is
defective and must be replaced. Perform
->Repair Procedures 3.2.6 . After replace-
ment of motor drive, test the unit for proper
functions and calibration.
Remove the cover from the load cell.
Check the operation if any of the Internal
drive motor. The cover has to be con-
nected to the base housing and power
applied. Cycle the internal calibration on
and off several times, if the motor drive
fails to respond, it is defective and must be
replaced. Perform ->Repair Procedures
3.2.8 . After replacement of motor drive,
test the unit for proper functions and cali-
bration.
TABLE 2-6 BALANCE CANNOT CALIBRATE INTERNALLY- AV MODELS ONLY.