EN-8 

 

 Analyzer and Bench Meter

 

2.  Installation 

2.1  Install the stand-alone electrode holder 

1.  Attach the magnetic part on the bottom of the meter to the magnetic part on the 

stand-alone electrode holder to finish installation. (Both marked in red 

demonstrated below.) 

 

 

2.  The two tension knobs can be adjusted by turning clockwise to tighten or turning 

counterclockwise to loosen. When they are loosened, you can move the holder up 

and down, forward and backward to the desired angle. Then you can tighten the 

knobs to fix the holder.  

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for a-AB33M1

Page 1: ...lti parameter Analyzer a AB33M1 Bench PH Meter a AB33PH a AB41PH Bench Conductivity Meter a AB33EC Instruction Manual www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply co...

Page 2: ...15 3 3 1 Measurement modes 15 3 3 2 PH measurement 16 3 3 3 mV and ORP measurement 16 3 3 4 Temperature measurement 16 4 CONDUCTIVITY OPERATION A AB33M1 AND A AB33EC 17 4 1 CONDUCTIVITY SETUP 17 4 1 1...

Page 3: ...NCE 38 11 3 ELECTRODE MAINTENANCE 38 11 3 1 PH electrode maintenance 38 11 3 2 Conductivity electrode maintenance 38 11 4 CLEANING 39 12 TECHNICAL DATA 39 12 1 SPECIFICATIONS 39 12 2 DIMENSIONS 44 13...

Page 4: ...trument malfunctions and false results WARNING For a hazardous situation with medium risk possibly resulting in severe injuries or death if not avoided CAUTION For a hazardous situation with low risk...

Page 5: ...spark formation corrosion caused by the ingress of gases WARNING Electrical shock hazards exist within the housing The housing should only be opened by authorized and qualified personnel Remove all p...

Page 6: ...rogress 6 Pinter icon touch to print 7 Calibration icon touch to calibrate blinks when calibration is in progress 8 Menu icon touch to enter menu settings 9 PH electrode condition a AB33M1 a AB33PH an...

Page 7: ...mV ORP a AB33M1 a AB33PH and a AB41PH EC TDS SAL Salinity RDS Resistivity a AB33M1 a AB33EC 13 Second measurement channel a AB33M1 Note a AB33M1 supports dual channel measurement other models only su...

Page 8: ...tep Home Button a Press to turn on the meter b Press to start a new measurement c Press to exit from the current mode or menu and return to the measurement mode Down Button a When setting parameters p...

Page 9: ...finish installation Both marked in red demonstrated below 2 The two tension knobs can be adjusted by turning clockwise to tighten or turning counterclockwise to loosen When they are loosened you can...

Page 10: ...s storage strip located at the bottom of the holder and pass the cable through the strip 2 2 Install the power adapter Change the plug for the power adapter according to the graphic below if necessary...

Page 11: ...lectrode the COND socket a AB33PH a AB41PH There are two sockets for pH electrode the pH mV socket and the Temp Socket Note for 2 in 1 pH electrode you only need to connect it to the pH mV socket whil...

Page 12: ...h Setup 3 Press the button if you are using an a AB33M1 model Skip this step if you are using an a AB33PH or a AB41PH model 4 Touch Temperature 5 Touch Temp unit and then touch C or F to select temper...

Page 13: ...as follows a pH electrode preparation b Buffer preparation and pH electrode calibration c Sample preparation d pH measurement e Record measurement results or print f Rinse the pH electrode and proper...

Page 14: ...ay the calibration point pH value e g 7 00 with the temperature a Touch Accept to accept the calibration result b Touch Recalibration if you want to re calibrate and repeat the previous steps 7 If you...

Page 15: ...the instruction on the screen to rinse the electrode place it in a standard solution stir 5 seconds and wait for 30 seconds 6 Press the button when you finish the previous step to start calibration T...

Page 16: ...nter menu option 2 Touch Setup and then touch Mode 3 Touch Read type 4 Touch Timed Time interval 00 hour 00 minutes 05 seconds is displayed 5 Press the or button to switch between different numbers Wh...

Page 17: ...ccuracy we recommend to use either a built in or a separate temperature electrode If a temperature electrode is used ATC and the sample temperature are displayed If the meter does not detect a tempera...

Page 18: ...t is out of the range the screen will display indicating the electrode needs cleaning 4 1 2 Set calibration standard solution The six predefined standard solutions are 10 S cm 84 S cm 146 5 S cm 500 S...

Page 19: ...uctivity calibration Set the calibration standard solution and cell K value before performing the calibration Please refer to the above Set calibration standard solution and Set cell K value section f...

Page 20: ...on 4 3 Conductivity measurement To switch between Auto and Continuous Modes before measurement please refer to Auto or Continuous End Point Mode section for details To start measurement 1 Place the el...

Page 21: ...point and the measurement ends The display will show the measurement result Note Touch the icon to print the result if you connect the analyzer to a printer or a computer Press the button to start a n...

Page 22: ...fferent users 1 Touch on the screen 2 Press the or button to select between different user IDs Touch when you see the ID number you want to confirm your setting 3 The display shows the following Enter...

Page 23: ...lts from 0 to 999 If the record is full the oldest record will be automatically overwritten To enter records management 1 Touch on the screen and then press the or button to view each record 2 Touch o...

Page 24: ...AB33M1 and 12 88 mS cm Ref Temp 20 C 68 F 25 C 77 F Temp Coeff 0 0 10 0 the default is 2 0 TDS Factor 0 01 5 0 the default is 0 50 Temperature Temp unit C F Enter temp value 5 0 60 0 41 0 140 the def...

Page 25: ...ter analyzer will consider the result as stable Fast stability When the difference between the maximum value and the minimum value of the measurement result within 4 seconds is less than 0 5pH or 0 6m...

Page 26: ...EC 146 5 S cm 500uS cm 1413uS cm a AB33M1 12 88mS cm Note please refer to Set calibration standard solution section for details of the standard solution setting Reference temperature Ref Temp 20 C 68...

Page 27: ...ution to measure temperature 3 Wait for the meter s temperature to become stable and check against the temperature measured by the NIST thermometer 4 If you need to enter the temperature measured by t...

Page 28: ...to purchase the ECS tester connector 2 Connect the ECS tester to the meter analyzer s pH mV socket as demonstrated in the following graphic 3 Touch the Menu icon to enter menu option 4 Touch i Steward...

Page 29: ...e reminder on the screen If all the above steps have been passed the meter analyzer will show a reminder that the self test has passed 7 Press the or button to return to the main menu If your meter an...

Page 30: ...button to confirm or press the button to cancel Parameter Reset Enter this sub menu to reset all the parameters under Setup menu To reset 1 Touch the Menu icon to enter menu option 2 Touch i Steward...

Page 31: ...12H Clock 24H Clock Date Format MM DD YY DD MM YY YY MM DD Adjustment Brightness High Medium Low Keypad Beep ON OFF Auto Shut Off OFF 15 30 About meter Language Set the language displayed for menus a...

Page 32: ...to enter menu option 2 Touch i Steward and then touch Adjustment 3 The first two digits of date on the upper left corner of the screen starts to blink 4 Press the or button to adjust the date Touch Ne...

Page 33: ...r analyzer will shut down after 15 minutes of no operation 30 the meter analyzer will shut down after 30 minutes of no operation Note during auto or continuous measuring modes the meter analyzer will...

Page 34: ...pH 11 mV Value XX X 12 mV Unit mV 13 Temperature Value XX X 14 Temperature Unit C F 15 Slope Value XXX X 16 Slope Unit 17 Offset Value XXX X 18 Offset Unit mV 19 Calibration ID a AB41PH 1 10 Table 9 2...

Page 35: ...ATC MTC 14 Calibration Type Manual 15 Cell Constant X X 16 Calibration ID 1 Table 9 4 PH measurement Number Mode Printing Content pH mV 1 Meter Model AB33PH AB33M1 AB41PH 2 Serial Number XXXXXXXXXX 3...

Page 36: ...odel AB33EC AB33M1 2 Serial Number XXXXXXXXXX 3 Software Revision X XX 4 User ID 0 9 5 Date Time 2020 9 25 14 34 6 Sample ID 000 999 7 Channel Cond 8 Mode Cond TDS SALT RES 9 Conductivity Value TDS Va...

Page 37: ...nd 7 The third row the above item number 8 9 10 11 and 12 The fourth row the above item number 13 14 and 15 The fifth row the above item number 16 17 18 and 19 The sixth row the above item number 20 a...

Page 38: ...the buffer has not been used more than once Error 7 a AB33M1 Cell K value out of range Make sure the Cell K value set in the menu is in compliance with the cell K value of the electrode you use The e...

Page 39: ...trode response becomes sluggish or the slope is not acceptable try the following Soak the electrode in 0 1M HCl for at least 8 hours For fat or oil contaminant degrease the membrane with cotton wool s...

Page 40: ...ply 12VDC 0 5A For use with certified or approved power supply which must have a SELV and limited energy output Voltage fluctuations Mains supply voltage fluctuations up to 10 of the nominal voltage O...

Page 41: ...s data points with time and date stamp Datalog for Calibration Last calibration 10 sets Keypad Capacitive touch Output Connect to PC via RS232 and USB Connect to Printer via RS232 pH Electrode Input B...

Page 42: ...r Groups 3 5 ORP RmV Measuring Range 2000 0 mV 2000 0mV Resolution 1 mV 0 1 mV Accuracy 1 mV 0 3 mV Units mV RmV Temperature Measuring Range 5 0 to 110 C 10 0 to 125 0 C 23 F to 230 F 14 F to 257 F Re...

Page 43: ...200 g L Resolution 0 01 mg L minimum auto ranging Accuracy 0 5 Reading 2 LSD TDS Factor Range Linear 0 04 to 10 00 default 0 5 Resistivity Measuring range 1 to 100 M cm Resolution 0 01 cm auto rangin...

Page 44: ...001 to 10 000 cm 1 Temperature Compensation Linear 0 to 10 0 C off TDS Measuring range 0 1mg L to 199 9 g L Resolution 0 01 mg L minimum auto ranging Accuracy 0 5 Reading 2 LSD TDS Factor Range Linear...

Page 45: ...EN 44 Analyzer and Bench Meter 12 2 Dimensions Figure 12 5 Dimensions 210 mm 142 mm 51 mm www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 46: ...uffer group JJG standard Temp C pH4 pH6 86 pH9 18 pH12 45 5 3 99 6 95 9 4 12 73 10 4 6 93 9 33 12 67 15 4 6 89 9 27 12 59 20 4 6 88 9 23 12 52 25 4 6 86 9 18 12 45 30 4 01 6 85 9 14 12 36 35 4 02 6 84...

Page 47: ...454 30 1 685 4 015 6 853 9 144 12 289 35 1 691 4 026 6 845 9 11 12 133 40 1 697 4 036 6 837 9 076 11 984 45 1 704 4 049 6 834 9 046 11 841 50 1 712 4 064 6 833 9 018 11 705 13 2 Conductivity standard...

Page 48: ...SO4 10 1 28 HF 1 5 7 2 coefficients of conductivity standards for a calculation to a reference temperature of 25 C Table 13 7 coefficients of conductivity standards Standard Measurement temp 15 C Meas...

Page 49: ...f Conformity is available online at This product complies with the EU Directive 2012 19 EU WEEE Please dispose of this product in accordance with local regulations at the collecting point specified fo...

Page 50: ...uipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause h...

Reviews: