4. Conductivity Operation (a-AB33M1 and a-
AB33EC)
This chapter is for a-AB33M1 and a-AB33EC models which support conductivity
measurement.
4.1 Conductivity setup
4.1.1 Set cell K value
The cell K value is usually recorded in the factory inspection report or marked on the
electrode cable.
To set the Cell K value:
1. Touch the
Menu
icon
to enter menu option.
2. Touch
Setup
, and then touch
Conductivity
.
3. Touch
Cell K
.
4. Press the
or
button to adjust the value. Touch the value on the screen to
confirm your setting.
After calibration, if the result is within ± 20% of the cell K value you set, the screen will
display
, indicating the electrode is in good condition. If the result is out of the range,
the screen will display
, indicating the electrode needs cleaning.
4.1.2 Set calibration standard solution
The six predefined standard solutions are 10 μS/cm, 84 μS/cm, 146.5 μS/cm, 500
μS/cm, 1413 μS/cm and 12.88 mS/cm.
To set the standard solution:
1. Touch the
Menu
icon
to enter menu option.
2. Touch
Setup
, and then touch
Conductivity
.
3. Touch
Std.Solution
, and then touch the standard solution you want to confirm
your setting. Press the
button to turn to the next page if the solution is not in
the first page.
4.1.3 Set reference temperature
To set the reference temperature:
1. Touch the
Menu
icon
to enter menu option.
2. Touch
Setup
, and then touch
Conductivity
.
3. Touch
Ref.Temp.
.
4. Touch the reference temperature you want to confirm your setting.
Note
: The calculation formula from °F to °C is: °C = 5/9 (°F - 32).