Analyzer and Bench Meter
EN-25
JJG
DIN
EU
Merck
Note
: please refer to
Select a predefined buffer group
section for details.
Conductivity (a-AB33M1, a-AB33EC)
Cell K
0.01~10.0
The default value is 0.1 for a-AB33EC model and 1.75 for a-AB33M1 model.
The cell K value of the compatible conductivity electrode AquaSeareacher series is as
follows:
Series
Cell K value
STCON7
0.08-0.1
STCON3
1.5-2.0
Note
: please refer to
Set cell K value
section for details of the cell K value setting.
Standard solution (Std.Solution)
10uS/cm
84uS/cm
(a-AB33EC)
146.5 μS/cm,
500uS/cm
1413uS/cm
(a-AB33M1)
12.88mS/cm
Note
: please refer to
Set calibration standard solution
section for details of the
standard solution setting.
Reference temperature (Ref.Temp.)
20°C (68°F)
25°C (77°F)
Note
: please refer to
Set reference temperature
section for details of the reference
temperature setting.
Temperature coefficient (Temp.Coeff.
)
0.0~10.0%/
℃
(the default is
2.0%/
℃
)
Note
: please refer to
Set Temperature coefficient
section for details of the
temperature coefficient setting.
TDS Factor
: 0.01~5.0 (the default is
0.50
)
Note
: please refer to
Set TDS factor
section for details of the TDS factor setting.
Temperature
Temperature unit
°C
°F