Specifications

Maximum Temperature: 500°F (260°C)

Maximum Pressure: 2,000 psi (13.8 MPa)

For tests above 200°F, refer to the chart below for the appropriate pres-

sure.

Do not use nitrous oxide cartridges as pressure sources for high-

temperature, high-pressure (HTHP) aging. Under high temperature and 

pressure, nitrous oxide can detonate in the presence of grease, oil, or 

carbonaceous materials. Nitrous oxide cartridges are to be used only 

for Garrett Gas Train Carbonate Analysis.

Carbon dioxide and nitrous oxide cartridges are pressurized to approxi-

mately 900 psi at 1 atmosphere (sea level). Therefore, they should never 

be transported by airplane without proper packing because cabin de-

pressurization may cause an explosion.

If the aging cells are going to be rolled in a roller oven during a test,  

install o-rings on the outer perimeter on the top and bottom of the cells. 

Failure to do so can damage the rollers in the oven. Teflon (#175-46), 

Viton

®

 (#175-47), and Buna N (#175-54) o-rings are available.

Cell Corrosion

Test fluids under high tempreature and pressure can corrode the cell body 

and caps. Carefully inspect the cell body and calls for corrosion before and 

after each test.

Some materials are more susceptible to corrosion than other. Also, some 

fluids and additives are more corrosive than others. OFITE offers a variety of 

cell materials for different levels of corrosion resistance and cost.

Mud Volume and Pressure for 

High-Temperature Aging

Aging 

Temp. 

(°F / °C)

Water 

Vapor 

Pressure 

(psi)

Coefficient 

of

Expansion 

of Water

Suggested 

Applied 

Pressure 

(psi / kPa)

Mud

Volume in

260 mL

Cell (mL)

Volume 

with 

Teflon 

Liner

Mud 

Volume 

in 500 

mL Cell 

(mL)

Volume 

with 

Teflon 

Liner

212 / 100

14.7

1.04

25 / 172

225

130

450

326

250 / 121

30

1.06

50 / 345

225

130

450

326

300 / 149

67

1.09

100 / 690

200

116

425

308

350 / 176

135

1.12

150 / 1,034

200

116

400

289

400 / 204

247

1.16

250 / 1,724

-

-

375

271

450 / 232

423

1.20

300 / 2,069

-

-

375

253

500 / 260

680

1.27

375 / 2,586

-

-

325

235

Summary of Contents for 175-25

Page 1: ...Stainless Steel Grade 303 500 mL 175 50 Stainless Steel Grade 316 500 mL Instruction Manual Updated 6 14 2019 Ver 4 OFI Testing Equipment Inc 11302 Steeplecrest Dr Houston Texas 77065 U S A Tele 832 3...

Page 2: ...lecrest Dr Houston TX 77065 USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 1 Intro 2 Components 3 Parts and Accessories 4 Specifications 5 Operation 6 Maintenance 8 Warranty and Return Policy 9 T...

Page 3: ...perature testing up to 500 F 260 C For prolonged exposure to elevated salinity at high temperatures such as 20 000 mg L chlorides at 350 F 176 6 C cells con structed from premium metals such as Hastel...

Page 4: ...g Cell Body For Temperatures Below 400 F Qty 3 175 14 Set Screw Diameter Qty 3 175 15 Wrench for Set Screw 175 16 Valve Stem 175 47 O ring for Outside of Aging Cell Body Viton 90 Qty 2 175 30 SP Spare...

Page 5: ...17 Viton O ring for Valve Stem 175 09 2 Viton O ring for Inside of Aging Cell Body 175 47 Viton O ring for Outside of Aging Cell 175 62 Viton O ring for Teflon Liner Plug 175 63 Viton O ring for Teflo...

Page 6: ...e outer perimeter on the top and bottom of the cells Failure to do so can damage the rollers in the oven Teflon 175 46 Viton 175 47 and Buna N 175 54 o rings are available Cell Corrosion Test fluids u...

Page 7: ...ill be tested refer to the chart on page 5 Place the correct volume of fluid into the cell Be careful not to get fluid in the o ring groove or on the o ring Do not overfill the cell 3 Clean any spille...

Page 8: ...ell from the oven and allow it to air cool until the sample temperature is 300 F 149 C or less The cell may then be either air or water cooled The sample temperature must be equal to the ambient tempe...

Page 9: ...e to the threads on the outer cap and cell body Every three years yearly for highly corrosive conditions 1 Sand blast the metal components cell body inner cap outer cap 2 Visually inspect the cell bod...

Page 10: ...ny losses or damages that occur as a result of Improper installation or maintenance of the products Misuse Neglect Adjustment by non authorized sources Improper environment Excessive or inadequate hea...

Reviews: