background image

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

8

Maintenance

Regular maintenance will extend the useful life of the aging cells.

After every test:

1.  Thoroughly clean and dry all components including the cell body,   

inner cap, outer cap, valve stem, and o-rings.

2.  Inspect all o-rings for damage or wear. Discard any that are nicked, 

cracked, or distorted from their original shape. Replace them with new 

o-rings.

3.  Apply grease to the threads on the outer cap and cell body.

Every three years (yearly for highly corrosive conditions):

1.  Sand blast the metal components (cell body, inner cap, outer cap).

2.  Visually inspect the cell body and inner cap for pitting or other signs of 

corrosion. If any material loss is present, replace the damaged com-

ponent.

3.  Inspect the threads on the outer cap and cell body for burrs.

4.  Replace all o-rings and apply grease to the threads on the outer cap 

and cell body.

5.  Pressure test with water up to 4,000 psi.

OFITE offers aging cell recertification that includes all of the above plus a 

one-year certificate. Contact your sales representative for more information.

Summary of Contents for 175-25

Page 1: ...Stainless Steel Grade 303 500 mL 175 50 Stainless Steel Grade 316 500 mL Instruction Manual Updated 6 14 2019 Ver 4 OFI Testing Equipment Inc 11302 Steeplecrest Dr Houston Texas 77065 U S A Tele 832 3...

Page 2: ...lecrest Dr Houston TX 77065 USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 1 Intro 2 Components 3 Parts and Accessories 4 Specifications 5 Operation 6 Maintenance 8 Warranty and Return Policy 9 T...

Page 3: ...perature testing up to 500 F 260 C For prolonged exposure to elevated salinity at high temperatures such as 20 000 mg L chlorides at 350 F 176 6 C cells con structed from premium metals such as Hastel...

Page 4: ...g Cell Body For Temperatures Below 400 F Qty 3 175 14 Set Screw Diameter Qty 3 175 15 Wrench for Set Screw 175 16 Valve Stem 175 47 O ring for Outside of Aging Cell Body Viton 90 Qty 2 175 30 SP Spare...

Page 5: ...17 Viton O ring for Valve Stem 175 09 2 Viton O ring for Inside of Aging Cell Body 175 47 Viton O ring for Outside of Aging Cell 175 62 Viton O ring for Teflon Liner Plug 175 63 Viton O ring for Teflo...

Page 6: ...e outer perimeter on the top and bottom of the cells Failure to do so can damage the rollers in the oven Teflon 175 46 Viton 175 47 and Buna N 175 54 o rings are available Cell Corrosion Test fluids u...

Page 7: ...ill be tested refer to the chart on page 5 Place the correct volume of fluid into the cell Be careful not to get fluid in the o ring groove or on the o ring Do not overfill the cell 3 Clean any spille...

Page 8: ...ell from the oven and allow it to air cool until the sample temperature is 300 F 149 C or less The cell may then be either air or water cooled The sample temperature must be equal to the ambient tempe...

Page 9: ...e to the threads on the outer cap and cell body Every three years yearly for highly corrosive conditions 1 Sand blast the metal components cell body inner cap outer cap 2 Visually inspect the cell bod...

Page 10: ...ny losses or damages that occur as a result of Improper installation or maintenance of the products Misuse Neglect Adjustment by non authorized sources Improper environment Excessive or inadequate hea...

Reviews: