TUF-2000F User Manual
2
D is the pipe diameter
Tup is the time for the beam from upstream transducer to the downstream one
Tdown is the time for the beam from downstream transducer to the upstream one
Δ
T=Tup –Tdown
1.4 Optional transducer
TS
◆
-1
clamp-on type transducer, pipe size from DN15-100mm
TM
◆
-1
clamp-on type transducer, pipe size from DN50-1000mm
TL
◆
-1
clamp-on type transducer, pipe size from DN300-6000mm
1.5 Typical application
The wall-mounting flow meter can be applied to a wide range of pipe flow measurements. Applicable liquids
include pure liquids as well as liquid with small quantity of tiny particles.
Examples are:
★
Water (hot water, chilled water, city water, sea water, waste water, etc.);
Sewage with small particle conte
★
nt;
Oil (crude oil, lubricating oil, diesel oil, fuel oil, etc.);
★
Chemicals (alcohol, acids, etc.);
★
Plant effluent;
★
Beverage, liquid food;
★
Ultra
★
-pure liquids;
Solvents and other liquids
★
1.6 Product Identification
Each set of the flow meter has a unique product identification number or ESN (electronic serial number)
written into the software that can only be modified with a special tool by the manufacturer. In
case of any hardware failure, please provide this number which is located on menu window
M61
when
contacting the manufacturer.
1.7 Specifications
Items
Specifications
Accuracy
Better than ±1%
Repeatability
Better than 0.2%
Principle Transit-time
measuring principle
Measurement
Period
500ms
Main unit
Display
LCD with backlight,
display accumulated flow/heat, instantaneous
flow/heat, velocity, time etc.