OPM223/253
Operation
55
R262
Enter set point 1 (or
2)
pH 6.00
pH -2.00 ... 16.00
Relay assignment 1 and 2 for neutra
controller:
Rel1 = set point 1
Rel2 = set point 2
R263
Enter relay
switching capacity
K
p
1 (or K
p
2)
1.00
0.10 ... 20.00
Relay assignment 1 and 2 for neutra
controller:
Rel1 = Kp1
Rel2 = Kp2
R264
Enter integral action
time T
n
1 (or T
n
2)
(0.0 = no
I-component)
0.0 min
0.0 ... 999.9 min
Relay assignment 1 and 2 for neutra
controller:
Rel1 = Tn1
Rel2 = Tn2
R265
Enter derivative
action time T
v
1 (or
T
v
2)
(0.0 = no
D-component)
0.0 min
0.0 ... 999.9 min
Relay assignment 1 and 2 for neutra
controller:
Rel1 = Tv1
Rel2 = Tv2
R266
Select pulse length
or pulse frequency
Len = pulse length
Freq = pulse frequency
Curr = current output 2
Pulse length e.g. for solenoid valve,
pulse frequency e.g. for solenoid dosing
pump, see "Actuating signal outputs"
section.
Curr = current output 2 can only be
selected if O2 = Contr.
R267
Enter pulse interval
10.0 s
0.5 ... 999.9 s
This field only appears if pulse length is
selected in R266.
If pulse frequency is selected, R267 is
skipped and entries continue with R268.
R268
Enter maximum
pulse frequency of
the adjuster
120 min
-1
60 ... 180 min
-1
This field only appears if pulse
frequency is selected in R266.
If pulse length is selected, R268 is
skipped and entries continue with R269.
R269
Enter minimum
switch-on time t
ON
0.3 s
0.1 ... 5.0 s
This field only appears if pulse length is
selected in R266.
R2610 Enter process type
Batch
Inlne
Batch = discontinuous process
Inlne = continuous process
There is no further dosing in the setting
range in batch mode. The I-component
is decreased.
Dosing continues in the setting range in
inline mode. The I-component is
effective.
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
R262
pH
Setpoint1
6
.
00
R263
Kp Rel1
1
.
00
R264
min
Tn Rel1
0
.
0
R265
min
TV Rel1
0
.
0
R266
Oper
.
Mode
len
R267
s
Puls Per
.
10
.
0
R268
1
/
min
Max
.
PFreq
120
R269
s
Min
.
PTime
0
.
3
R2610
%
Proc
.
Type
Batch