background image

                       Odyssey

®

 Xtreem  

                                                 Single Soil Moisture Logger 

             

User Manual

In the calibration tab select the soil type as Soil Test for all levels and press the submit button. We 
suggest the logging interval is set to 1 minute for this test.
Place a 250mm diameter cylinder ( a 20 litre bucket with the bottom cut out ) around the logger and 
fill with water. The water will slowly empty from the bucket into the soil. This may have to be repeated 
several times depending on the soil type and the depth of the bottom sensor.
Using the Xpert portal select the chart page for this logger. It can be clearly seen that the moisture 
value increases soon after the water is applied around the logger. The moisture value should rise to a
peak value then drop back a small amount and level out. The measured field capacity from the chart 
page must be multiplied by 100 when entering it into the 'Field Cap' box on the Calibration page.
Find a soil type on the Calibration page where the field capacity is similar to the measured field 
capacity ( x 100 ) and the drypoint value for this soil type should be the drypoint needed. Note the 
values entered on this calibration page do not affect the factory calibration of the logger, just the 
displayed and exported data. Choose the Custom Soil type from the drop down box and enter the 
drypoint value and field capacity value. 

Volumetric Measurement

There are three more parameters required for Xpert to calculate the parameters to give volumetric 
soil moisture percentage.
These measurements are normally carried out by a qualified technician following a test Standard like 
ASTM D2216 there the following is only an overview to aid in the understanding of the process.
Using a soil corer collect a saturated soil sample and place this sample into a plastic bag. The 
volume of soil the corer collects is the 

Saturated Volume

Accurately weigh the soil sample, this is the 

Saturated Soil weight

 . 

Dry the soil, this is normally done in an oven over several days. 
Accurately weigh the dry soil, this is the 

Dry Soil weight.

Calculations

For either Percentage of Saturation or Volumetric Percentage the calculation is the same.
The slope and offset parameters are used as follows to calculate soil moisture.

Soil Moisture (%) = 

((

Logger Reading

(

%

)

100

×

Slope

)+

Offset

100

For a logger reading of 60.00% with calculated scale of 1.4815 and an offset of -0.3704.

((

60.00

100

×

1.4815

)−

0.3704

100

= 51.85%

DATAFLOW SYSTEMS LTD

Page 6

www.Odysseydatarecording.com

 

            V 03/21

Summary of Contents for Xtreem

Page 1: ...Odyssey Xtreem Single Soil Moisture Logger User Manual Xtreem Single Soil Moisture Logger User Diagram DATAFLOW SYSTEMS LTD Page 1 www Odysseydatarecording com V 03 21...

Page 2: ...Configuration 4 Calibration 4 Transfer Configuration to Logger 5 Soil Types 5 Volumetric Measurement 6 Calculations 6 Collecting Data 7 Viewing Data 7 ICON pop up 7 Selected Logger Box 7 Latest Infor...

Page 3: ...sor cable must be plugged into the logger The connector is keyed so once this is lined up push the connector in as far as it goes and then rotate the top of the connector until it locks into place The...

Page 4: ...oad the full Odyssey Xpert user manual from Xpert web portal Logger Configuration 1 Select the Logger Config Tab Enter a Logger Name that will be used to Identify the Logger from now on The serial num...

Page 5: ...s than 10 minutes through the tablet phone running the Xtract Application To apply this immediately press the SERVER SYNC button in the Settings screen Once the button has been pressed the time below...

Page 6: ...calibration page do not affect the factory calibration of the logger just the displayed and exported data Choose the Custom Soil type from the drop down box and enter the drypoint value and field capa...

Page 7: ...then select a logger to be configured An Icon for the logger will display on the Map Click on the icon It may be necessary to Zoom the Map out to find the Logger ICON pop up This pop up shows brief da...

Page 8: ...rding at least 60000 records The number of recording days before the memory is full can be calculated using the following formulae If continuous logging is enabled the oldest will overwritten first If...

Page 9: ...er Note the unit will not be damaged if the batteries are inserted incorrectly Before replacing the cap check the O Ring seal is present clean and free from dirt If required apply some silicon grease...

Page 10: ...diation exposure limits set forth for an uncontrolled environment This equipment should be installed and operated with a minimum distance of 20cm between the radiator and your body Japan MIC The Xtree...

Reviews: