EX Series Operating Manual
6-7. OVP USE
▌Enable or disable the "Over Voltage Protection" function.
▌In case of a like inductivity DUT, motor, big capacity DUT and etc, occurs "OVP trip" frequently
so if it is not convenient to use the power supply, you can set disable of "OVP USE" mode.
▌If disable the "OVP USE", the power supply doesn't cut the DC source when voltage level is
over OVP setting level.
▌F
ti
▌Function
> DISABLE
Disable "OVP USE" mode.
> ENABLE
Enable "OVP USE" mode.
OVP USE
▌Power switch on in a pressing "FACTORY" key.
▌Power switch ON
▌When appearing "1. user-mem clear" message, release "FACTORY" key.
In a pressing
▌Enter to "5. OVP USE" mode by using cursor key.
▌In other to set, press the "FACTORY" key one more time.
LCD Display
Release
enable
▌Set the "enable" mode by using cursor key.
LCD Display
▌To finish the "OVP USE" mode, press the "FACTORY" key one more time.
LCD Display
disable
done
≫ Related remote interface command
FACT:OVP {DIS|ENA}
FACT:OVP?
Return Value(1, 0)
Apply:
FACT:OVP?
Return value : 0
"OVP USE" mode is disable state
FACT:OVP ENA
"OVP USE" mode changed to
enable state
- 79 -