MNL-1009 REV A
15
5.
Uncheck the
Value
box to open the shutter.
External Triggering Options
You can trigger the QE
Pro
using a variety of External Triggering options through the 30-pin Accessory Connector on the
spectrometer. See the External Triggering Options document located at
https://www.oceaninsight.com/support/documents-
manuals/manuals-operating-instrs/
. The document contains instructions for configuring External Triggering options for the QE
Pro
.
Changing the Slit
QE
Pro
has an interchangeable slit feature allow you to move rapidly between different measurement types without changing your
spectrometer.
For instructions on how to change the slit on your spectrometer, visit
Ocean Insight recommends that a calibration always be performed after changing the slit on a spectrometer as it will optimize the
performance of the device.
NOTE
Make sure the ALT checkbox for the GPIO (Index 4) is not selected, which would prevent the shutter from triggering.
The activation time of the internal shutter is 11 ms.