TABLE RESET PROCEDURE - HAND CONTROL
DIRECTIONS FOR USE
1. Press & hold the 2nd set of buttons simultaneously -
wait 5-6 seconds for retract to begin.
2. Hold until the columns are fully retracted then continue
holding for 5 seconds after full retraction.
3. After the table is reset, follow the “Set Zero Height”
procedure on the next page.
FLAT TOP OR BACKREST
*FLAT TOP OR BACKREST W/ABC
SALON TOP
SALON TOP W/ABC
*Feel around to find the buttons underneath the overlay
MEM1
MEM2
MEM1
MEM2
1. Locate the Control Box and detach the strain relief by removing
the screw holding it in place with a 2.5mm Allen wrench.
2. Unplug the foot control from the 3rd control port in the TC8
Control box.
3. Plug it into the 1st control port.
4. Press & hold the 2nd set of buttons simultaneously - wait 5-6
seconds for retract to begin.
5. Hold until the columns are fully retracted then continue holding
for 5 seconds after full retraction.
6. Return the foot control plug to the 3rd control port and replace
the strain relief.
7. After the table is reset, follow the “Set Zero Height” procedure on
the next page.
TABLE RESET PROCEDURE - FOOT CONTROL
Port 3
Port 1