4.4.2 

Setting the TDS Conversion Factor (P4.4) 

The factory default setting for TDS conversion factor is 0.5. If your solution has a different TDS factor, 
you can improve accuracy by setting the TDS factor prior to calibration. 
1. Press 

SETUP. 

2. Press 

MI/

 or MR/

, go to P4.0. 

3. 

Press ENTER a five times to scroll until you view P4.4 (tdS) 

4. 

Press ENTER. The upper display shows a value and the lower shows “tdS”. 

5. Press MI/

 or MR/

 to select your calculated TDS conversion factor. 

6. Press the ENTER to confirm. 
7. Press the CAL/MEAS key to return to measurement mode. 

4.4.3 

Calibration with Conductivity Standard and TDS factor 

After setting the correct TDS Factor, you can calibrate in the TDS mode. 
1. 

Calibrate with conductivity standards as described in 4.4 

2. 

Set the TDS factor as mentioned above (if not yet done). 

5 MEASUREMENT 

5.1 Temperature 

Compensation 

This meter can take measurements with automatic (ATC) or manual (MTC) temperature 
compensation. Factory default is ATC on. For ATC, attach the conductivity/TDS electrode to the 
meter. The ATC indicator will light on the LCD. 
If the conductivity/TDS electrode is not properly attached to the meter or it has been damaged, the 
ATC indicator will blink and the temperature display will show “Ur”. If the ATC indicator does not light 
and the temperature display shows a reading, the meter might be in MTC mode. 

5.1.1 

Selecting Temperature Compensation mode (P4.3) 

1. Press 

SETUP 

2. Press 

MI/

 or MR/

, go to P4.0. 

3. 

Press ENTER three times, to go to P4.3  The upper display shows “ATC” and the lower display 
shows “P4.1”. 

4. Press 

ENTER 

5. Press 

MI/

 or MR/

 to select: “YES”, ATC is on 

or

 “NO”, ATC is off. 

6. 

Press ENTER to confirm selection. 

7. 

Press CAL/MEAS to return to measurement mode. 

Note:

 MTC should only used if ATC fails. For non-compensated measurements, set the coefficient to 

0.0%.

 

5.1.2 

Setting manual temperature compensation value 

MTC enables you to enter the temperature value of your sample into the meter. This is the value at 
which the readings will be temperature compensated. You can select an value between 0 and 100 °C 
(32 and 212 °F). Default value is 25 °C. 
1. 

Set temperature mode to MTC (p 4.1.1) and go through step 1 to 7. 

2. 

Press CAL/MEAS. The CAL indicator will appear above the primary display. 

3. 

Press MODE. The primary display shows the current setting, the secondary display shows 25 °C 
(77 °F) or its last set value. 

4. Press 

MI/

 or MR/

 to offset the temperature to your preferred settings. 

5. 

Press ENTER to confirm 

Note: 

To exit this program without confirming the MTC value, press CAL/MEAS. 

5.2 Taking 

Measurements 

1. 

Rinse the electrode with de-ionized or distilled water before use. Shake or air dry. To avoid 
contamination or dilution of your sample, rinse electrode with a small volume of your sample 
liquid. 

2. 

Press ON to switch on meter. 

 

6

Summary of Contents for CON 400

Page 1: ...Instruction Manual CON 400 and CON 410 Waterproof Hand held Conductivity TDS Meter ENGLISH Eutech Instruments Pte Ltd Oakton Instruments All rights reserved Version 1 1 ML...

Page 2: ...EADY and Auto HOLD function P4 1 7 5 5 Ranging settings and options 7 5 5 1 Selecting manual or auto ranging 7 6 MEMORY FUNCTION 7 6 1 Memory Input 7 6 2 Memory Recall 8 6 3 Memory Clear 8 7 OTHER FUN...

Page 3: ...libration sol 17 K Cell Constant 18 ON OFF indicator 19 HOLD Hold 20 READY Ready 2 2 Keypad Some button has several functions depending on its mode of operation KEY Function ON OFF ON OFF Powers on an...

Page 4: ...yond upper steel band Note DO NOT remove the protective electrode guard during measurement and calibration as it may affect your readings Note We recommend that you do not submerge the electrode above...

Page 5: ...and return to the measurement mode Note To exit without confirming the temperature calibration value press CAL MEAS Note If the ATC does not light up see SETUP menu Program P1 3 to switch it on Note S...

Page 6: ...rature display shows a reading the meter might be in MTC mode 5 1 1 Selecting Temperature Compensation mode P4 3 1 Press SETUP 2 Press MI or MR go to P4 0 3 Press ENTER three times to go to P4 3 The u...

Page 7: ...MR to select the configuration you require left display a ON activates the READY function b OFF deactivates the READY function c ON and HOLD activates the Auto HOLD function 5 Press ENTER to confirm...

Page 8: ...ress MI or MR to select between C and F 5 Press ENTER to confirm 6 Press CAL MEAS to return to measurement mode 7 2 Viewing calibration data P2 0 1 Press SETUP 2 Press MI or MR until you view paramete...

Page 9: ...you view parameter P5 0 3 Press ENTER three times to select P5 2 display shows t nr 4 Press ENTER 5 Press MI or MR to select either the normalization temperature 6 Press ENTER to confirm and return t...

Page 10: ...play Memory checksum error Hardware failure Err 3 in primary display A D converter error Hardware failure Err 4 in primary display Keypad error One or more keys on the keypad are stuck or fault in key...