8

2. 

If necessary, measure the next sample solution and press MI/

 key to input the next data into 

the memory. 

Note:

 If the memory is full, the first value stored will be overwritten. 

6.2 Memory 

Recall 

1. Press 

MR/

 once to retrieve the last reading stored. 

2. 

Press ENTER to recall the reading 

3. 

Press ENTER to view date and time 

4. 

Press ENTER to return. The display automatically moves to the next memory location screen. 

5. Press 

MI/

 or MR/

 to scroll to next or ENTER / RANGE to the previous memory location. 

6. 

Press MEAS to exit Memory Recall. 

Note: 

Readings stored in memory are retained even if the unit is turned off. 

6.3 Memory 

Clear 

1. Press 

SETUP. 

2. Press 

MI/

 or MR/

 until P1.0. 

3. Press 

ENTER. 

4. Press 

MI/

 or MR/

 to toggle between NO and YES. “NO” retains current memory, “YES” clears 

all

 memory. 

5. 

Press ENTER to confirm selection 

6. 

Press CAL/MEAS to return to measurement mode. 

7 OTHER 

FUNCTIONS 

7.1 

Selection of °C or °F (P4.2)

 

1. Press 

SETUP. 

2. Press 

MI/

 or MR/

 until you view parameter P 4.0. 

3. 

Press ENTER (2x), P4.2. 

4. Press 

MI/

 or MR/

 to select between °C and °F. 

5. 

Press ENTER to confirm 

6. 

Press CAL/MEAS to return to measurement mode. 

7.2 

Viewing calibration data (P2.0) 

1. Press 

SETUP. 

2. Press 

MI/

 or MR/

 until you view parameter P2.0 . 

3. 

Press HOLD/ENTER repeatedly to view the previous calibration data on each of the five 
measuring ranges. 

4. 

When you have scrolled through all calibration data, you will automatically return to the subgroup 
menu. 

5. 

Press CAL/MEAS key to return to the measurement mode. 

Note:

 If there is no calibration data at a particular point, the primary display will show “----“. 

Note:

 CON 410 only: Entering Set Up from Conductivity mode, data will be in 

µ

S or mS, from TDS 

mode, data will be in ppm or ppt. 

7.3 

Viewing Electrode Data (P3.0) 

1. Press 

SETUP. 

2. Press 

MI/

 or MR/

 until you view parameter P3.0. 

3. 

Press ENTER repeatedly to view the effective cell constant for each range. 

4. 

When you have scrolled through all the electrode data, you will automatically return to the 
subgroup menu. 

5. 

Press CAL/MEAS to return to measurement mode. 

7.4 

Reset to factory defaults (P8.0) 

Resets all settings to the factory default, except for clock settings. 
1. Press 

SETUP. 

2. Press 

MI/

 or MR/

 until you view parameter P8.0. 

3. Press 

ENTER. 

Summary of Contents for CON 400

Page 1: ...Instruction Manual CON 400 and CON 410 Waterproof Hand held Conductivity TDS Meter ENGLISH Eutech Instruments Pte Ltd Oakton Instruments All rights reserved Version 1 1 ML...

Page 2: ...EADY and Auto HOLD function P4 1 7 5 5 Ranging settings and options 7 5 5 1 Selecting manual or auto ranging 7 6 MEMORY FUNCTION 7 6 1 Memory Input 7 6 2 Memory Recall 8 6 3 Memory Clear 8 7 OTHER FUN...

Page 3: ...libration sol 17 K Cell Constant 18 ON OFF indicator 19 HOLD Hold 20 READY Ready 2 2 Keypad Some button has several functions depending on its mode of operation KEY Function ON OFF ON OFF Powers on an...

Page 4: ...yond upper steel band Note DO NOT remove the protective electrode guard during measurement and calibration as it may affect your readings Note We recommend that you do not submerge the electrode above...

Page 5: ...and return to the measurement mode Note To exit without confirming the temperature calibration value press CAL MEAS Note If the ATC does not light up see SETUP menu Program P1 3 to switch it on Note S...

Page 6: ...rature display shows a reading the meter might be in MTC mode 5 1 1 Selecting Temperature Compensation mode P4 3 1 Press SETUP 2 Press MI or MR go to P4 0 3 Press ENTER three times to go to P4 3 The u...

Page 7: ...MR to select the configuration you require left display a ON activates the READY function b OFF deactivates the READY function c ON and HOLD activates the Auto HOLD function 5 Press ENTER to confirm...

Page 8: ...ress MI or MR to select between C and F 5 Press ENTER to confirm 6 Press CAL MEAS to return to measurement mode 7 2 Viewing calibration data P2 0 1 Press SETUP 2 Press MI or MR until you view paramete...

Page 9: ...you view parameter P5 0 3 Press ENTER three times to select P5 2 display shows t nr 4 Press ENTER 5 Press MI or MR to select either the normalization temperature 6 Press ENTER to confirm and return t...

Page 10: ...play Memory checksum error Hardware failure Err 3 in primary display A D converter error Hardware failure Err 4 in primary display Keypad error One or more keys on the keypad are stuck or fault in key...